Precautions for using a multifunctional aperture analyzer
Date: 2016-03-16Read: 15
I'm going to talk to everyone todayMulti functional aperture analyzerPrecautions for use.
  
1. The exterior of the multifunctional aperture analyzer should be cleaned at all times, at least once a week. Generally, a soft bristled brush, velvet cloth, or suede can be used to remove dust from the balance. When cleaning, be careful not to touch the balance parts directly with your hands to prevent moisture from causing metal oxidation and volume changes on the parts. Therefore, fine yarn gloves or extremely thin rubber gloves should be worn, and the metal smooth stripes should be followed to avoid damaging the smoothness of the parts. To avoid the retention of harmful substances, the bottom seat should be cleaned immediately after each weighing. The working edge of the agate blade on the crossbeam should be kept highly clean, and suede is often used to slide back and forth along its edge, using slow cleaning. The agate surface of the middle blade support and edge blade pad, as well as the agate bearings of each part, should also be cleaned with suede. After cleaning the wall of the damper with a soft bristled brush and suede, use a 20-30 times magnifying glass to observe if there are still small substances present.
  
2. Multi functional aperture analyzer cleaning detection tube: Long term use of the instrument will cause the background count to exceed the range. Specific solution: Regularly use a syringe to extract cleaning agent and insert it into the detection tube, extract or remove the cleaning solution, select the dirt around the detection tube, and use a 100X microscope to check for foreign objects in the detection hole. Be careful when disassembling each flushing component to prevent human damage to the machine.
  
  
3. Regularly inspect and clean the filter screen. Every week or every 300 specimens are tested, the filter screen is inspected and cleaned, and dust is removed from the filter screen using tweezers. The filter screen is rinsed with water, and if it is still not clean, a new filter screen needs to be replaced.
  
  
The above is the complete content of the precautions for using the multifunctional aperture analyzer. Welcome to continue visiting our website to learn more about the multifunctional aperture analyzer.