Application of Infrared Moisture Analyzer
Principle:
The hydrogen oxygen bond in water molecules absorbs near-infrared light of a specific wavelength (specific measurement band 1.94 µ m). At a specific wavelength, the reflected near-infrared energy is inversely proportional to the near-infrared energy absorbed by water molecules in the material. The moisture content of the tested material is calculated based on the amount of energy loss.
When water molecules are excited by a specific energy band, they become vibrational states, and the hydrogen oxygen bond between two hydrogen atoms bound in the water molecule and one oxygen atom will stretch, contract, or twist in other forms. The energy that can cause water molecules to vibrate covers specific bands throughout the entire electromagnetic spectrum. In different parts of the entire spectrum, some absorption bands are very strong, while others are very weak. In the near-infrared region of the spectrum, this band is particularly strong for water molecules, and the instrument is easier to achieve in terms of emitting, filtering, and receiving energy.
Near infrared measurement technology is a non-destructive, non-contact real-time measurement technique.

Application of Infrared Moisture Analyzer
feature:
Highly integrated electronic devices, SMT technology circuit boards, reliable and effective output interfaces reduce possible fault points, and ensure stable and reliable quality
Dynamic real-time diagnostic systemNo need for regular maintenance
Can be equipped with a built-in cooling system
Built in product temperature measurement system can be added
High sensitivity: none
Sensitivity to material color/formula: small, good applicability.
The filter is smaller than traditional brand designs and less affected by bandwidth.
The stainless steel shell is used to prevent corrosion and water vapor dust.
The power supply and digital signal output are separated from the host, avoiding the instability of traditional brand output signals caused by the internal temperature of the probe.
There are more ways to output signals, including two sets of RS485 outputs and two sets of 4-20mA analog signal outputs.
ISF-300Infrared moisture analyzerSpecifications:
Measurement range: 0-95%
Measurement accuracy: ± 0.2% - ± 0.01%. (Value)
Resolution: 0.001%.
光斑:直径60mm
Measurement speed: 0.0125 seconds
Calibration and Calibration: Quick Calibration, Pre Calibration, No Need for Daily Re Calibration
Filtering time: linear or exponential (0.1-180 seconds)
Analog output: Two sets of 4-20mA outputs
Communication modes: RS232, RS485, Modbus, TCP/IP
Calibration file: can store 31 calibration files
Working temperature: -10--80 ℃
Working distance: 150-400mm
Size: 130X149X277mm
Weight: 5KG
Working power supply: 24VDC/35VA
Equipment protection: Stainless steel casing (IP65 protection level)
Clean blowing interface: blowing port diameter 6mm, compressed air 0.25MPa
Partial on-site application cases:

Online moisture detection of sintered mixtures

Online moisture detection of sintering mixture and incoming raw materials

Ceramic raw material pipeline transportation online moisture detection power plant coal moisture online detection

Chemical raw material online moisture detection, grain biochemistry online moisture detection

Online moisture detection of cigarette paper, online detection of propylene glycol in tobacco leaves