Assemble thermocoupleProduct Application
Usually used in conjunction with display instruments, recording instruments, electronic computers, etc. Directly measure the temperature of liquid, vapor, gas media, and solid surfaces within the range of (0~1300) ℃ in various production processes.
Product Principle
The electrodes of a thermocouple are composed of two different conductor materials. When there is a temperature difference between the measuring end and the reference end, a thermoelectric potential is generated, and the working instrument displays the temperature value corresponding to the thermoelectric potential.
Assemble thermocouplecharacteristic
·Easy to assemble and replace
·Pressure spring type temperature sensing element with good anti vibration performance;
·Large temperature measurement range
·High mechanical strength and good pressure resistance
Main technical parameters
Product execution standards
IEC584
IEC1515
GB/T16839-1997
JB/T5582-91
Room temperature insulation resistance
The insulation resistance between the electrode and the outer sheath of the thermocouple is greater than 1000 Ω. m when the ambient temperature is 20 ± 15 ° C, the relative humidity is not greater than 80, and the test voltage is 500 ± 50V (DC).
Measurement range and tolerance
| model | Dividing number | Tolerance level | |||
| I | II | ||||
| tolerance value | Temperature measurement range ° C | tolerance value | Temperature measurement range ° C | ||
| WRN | K | ±1.5°C | -40~+375 | ±2.5 °C | -40~+333 |
| ±0.004ltl | 375~1000 | ±0.0075 liters | 333~1200 | ||
| WRM | N | ±1.5°C | -40~+375 | ±2.5°C | -40~+333 |
| ±0.004 liters | 375~1000 | ±0.0075 liters | 333~1200 | ||
| WRE | E | ±1.5°C | -40~+375 | ±1.5°C | -40~+333 |
| ±0.004 liters | 375~800 | ±0.004 liters | 333~900 | ||
| WRF | J | ±1.5°C | -40~+375 | ±1.5°C | -40~+333 |
| ±0.004 liters | 375~750 | ±0.004 liters | 333~750 | ||
| WRC | T | ±1.5°C | -40~+125 | ±1°C | -40~+133 |
| ±0.004 liters | 125~350 | ±0.0075 liters | 133~350 | ||
Naming method for prefabricated thermocouples