CPS11D-7BA21 PH Electrode

CPS11D-7BA21 PH electrode is used in conjunction with an E+H meter
Electrode cleaning
If the electrode is not mechanically damaged and there are no damaged parts,
You can regenerate according to the following instructions
Salt deposition: Immerse the electrode in 0.1 M HCl hydrochloric acid solution for 5 minutes,
Dissolve the sediment and then immerse it in a 0.1 M NaOH solution
5 minutes, then rinse with plenty of deionized water
Lubricating grease/oil: Clean the electrode foam with detergent and water. Rinse the end of the electrode with deionized water
Reference electrode salt bridge blockage: Heat the diluted KCL solution to 60-80 º C,
Soak the blocked part of the electrode in a hot solution for 10 minutes. Then put it in the cool one
Cool the electrode in KCL solution
If the above methods cannot ensure a normal response time, replace the electrode immediately