Relay on-off capability life test device
JAY-3203L
1. Overview:
The JAY-3203 series relay connection and disconnection capability, electrical life testing equipment (hereinafter referred to as the testing equipment) is designed in accordance with the relevant requirements of GB14598.1-2001 and GB7261-2008 "Basic Test Methods for Relay Protection and Safety Automatic Equipment".
2. Main technical parameters of testing equipment:
2.1. Sample coil working voltage: DC stabilized voltage 3-20V, adjustable accuracy: ± 1% ± 2 characters.
2.2 LoadWorkstation: 1-3 workstations, working in a loop, sharing a set of loads
2.3.1. Rated current of sample contact: 1-10A
Test voltage (AC/DC): 5-250V, requires external supplyDisplay accuracy: ± 1%± 2 words
Load current: 1~60AAdjustable, current display accuracy: ± 1%± 2 words
Power factor: 0.3-1, display accuracy: ± 0.05
The rated current is 6 times the rated current within 100mS when the coil is powered on, and then decreases to 1 time the rated current afterwards
2.3.2. Rated current of sample contact: 1-20A
Test voltage (AC/DC): 5-250V, requires external supplyDisplay accuracy: ± 1%± 2 words
Load current: 1-120AAdjustable, current display accuracy: ± 1%± 2 words
Power factor: 0.3-1, display accuracy: ± 0.05
The rated current is 6 times the rated current within 100mS when the coil is powered on, and then decreases to 1 time the rated current afterwards
2.3.3. Rated current of sample contact: 1-30A
Test voltage (AC/DC): 5-250V, requires external supplyDisplay accuracy: ± 1%± 2 words
Load current: 1-180AAdjustable, current display accuracy: ± 1%± 2 words
Power factor: 0.3-1, display accuracy: ± 0.05
The rated current is 6 times the rated current within 100mS when the coil is powered on, and then decreases to 1 time the rated current afterwards
2.4. The load is composed of multiple levels of inductors and resistors connected in series, and the load can be converted from AC to DC;
Load resistance: adjustable from 0.025 to 300 Ω, with a minimum step value of 0.025 Ω
The load is switched by a knife switch
The load cooling method is forced air cooling, and its working system is a long-term continuous working system.
2.5 Operating frequency:
Power on time: 0-99.99sAccuracy: ± 1%;
Power outage time: 0-99.99sAccuracy: ± 1%.
There is an indicator light indicating the power on and off of the product
2.6. Test frequency range: 1-999999 times can be set; Accuracy: ± 1 time.
Maximum failure times: 1 to 9999 times can be set
Number of first occurrence failures: 1 to 999999 times can be recorded
When any test sample fails (normally open or normally closed), the number of failures shall be increased by one; When the set number of failures is reached, the system automatically disconnects the load and coil control voltage of the failed sample, and retains the count of the final test times at the faulty failure station.
2.7 The equipment has protective and alarm functions against phenomena such as fusion welding and short circuit of the test sample.
2.8. Install high temperature alarm, overcurrent protection, and AC leakage protection functions inside the load box.
2.9 Controller: Equipped with intelligent timers and counters, all data is saved in case of power failure
2.10. Power supply for equipment: AC220V 50/60Hz
2.11Equipment approximate dimensions: width 1800mm X depth 900mm X height 1800mm, weight approximately 700kg
Configure movable casters
2.12. Environmental conditions for use:
A、 Environmental temperature: 0 ℃~40 ℃, indoor type
B、 Relative humidity: Maximum relative humidity of 95% and no condensation, no corrosive gases or vapors, etc
C、 Altitude not exceeding 1000m
2.13. The approximate appearance of the equipment is as follows(Subject to physical object):