The Eppendorf Reference 2 pipette represents extraordinary precision, durability, reliable handling, and ensures user satisfaction, making it an ideal choice for handling precious liquid samples and applications with high precision requirements.
The Reference 2 pipette is an upgraded version of the classic Reference pipette, featuring a * new * * design, lighter weight, and less operating force. Additionally, the Reference series multi-channel pipette has been launched, once again pushing forward a new standard in the pipetting industry. As the * * product in the Eppendorf pipette series, the new Reference 2 pipette represents extraordinary precision, durability, reliable handling, and ensures user reference. The Reference 2 pipette is the * ideal choice for your liquid handling.
Product Features
1Single button technology, ergonomically designed, with minimal operating force; One step transfer, discharge, and removal of suction tips to reduce aerosols; No need to move the thumb, reducing repetitive strain injuries. There is tactile feedback before removing the suction head to prevent accidental operation
2Elastic suction nozzle improves the uniformity of liquid transfer and reduces the force required for assembling/disassembling the suction head
3The multi-channel pipette has an elastic suction nozzle function, which can be freely opened or closed
4Density adjustment function, which can adjust the density according to the properties of the solution, resulting in more accurate pipetting
5、Quick connection buckle, easy to remove the lower part, easy to clean or maintain
6High * * and high accuracy, providing reliable pipetting results
74-digit volume setting, digital zoom window for easy identification of volume
8Stainless steel frame, sturdy and durable
9Built in RFID technology, data can be read or added through Eppendorf TrackIT
Technical Specifications:
|
Product Model
|
volume
|
Inaccuracy
|
Not * * degree
|
|
|
|
± %
|
±μL/ mL
|
± %
|
±μL/ mL
|
|
0.1–2.5 μL, dark gray
|
0.1 μL
|
±48.0
|
±0.048
|
±12.0
|
±0.012
|
|
|
0.25 μL
|
±12.0
|
±0.03
|
±6.0
|
±0.015
|
|
|
1.25 μL
|
±2.5
|
±0.031
|
±1.5
|
±0.019
|
|
|
2.5 μL
|
±1.4
|
±0.035
|
±0.7
|
±0.018
|
|
0.5–10 μL, grey
|
0.5 μL
|
±8.0
|
±0.04
|
±5.0
|
±0.025
|
|
|
1 μL
|
±2.5
|
±0.025
|
±1.8
|
±0.018
|
|
|
5 μL
|
±1.5
|
±0.075
|
±0.8
|
±0.04
|
|
|
10 μL
|
±1.0
|
±0.10
|
±0.4
|
±0.04
|
|
2–20 μL, light gray
|
2 μL
|
±5.0
|
±0.10
|
±1.5
|
±0.03
|
|
|
10 μL
|
±1.2
|
±0.12
|
±0.6
|
±0.06
|
|
|
20 μL
|
±1.0
|
±0.20
|
±0.3
|
±0.06
|
|
2–20 μL, yellow
|
2 μL
|
±5.0
|
±0.10
|
±1.5
|
±0.03
|
|
|
10 μL
|
±1.2
|
±0.12
|
±0.6
|
±0.06
|
|
|
20 μL
|
±1.0
|
±0.2
|
±0.3
|
±0.06
|
|
10–100 μL, yellow
|
10 μL
|
±3.0
|
±0.3
|
±0.7
|
±0.07
|
|
|
50 μL
|
±1.0
|
±0.5
|
±0.3
|
±0.15
|
|
|
100 μL
|
±0.8
|
±0.8
|
±0.2
|
±0.2
|
|
20–200 μL, yellow
|
20 μL
|
±2.5
|
±0.5
|
±0.7
|
±0.14
|
|
|
100 μL
|
±1.0
|
±1.0
|
±0.3
|
±0.3
|
|
|
200 μL
|
±0.6
|
±1.2
|
±0.2
|
±0.4
|
|
30–300 μL, orange
|
30 μL
|
±2.5
|
±0.75
|
±0.7
|
±0.21
|
|
|
150 μL
|
±1.0
|
±1.5
|
±0.3
|
±0.45
|
|
|
300 μL
|
±0.6
|
±1.8
|
±0.2
|
±0.6
|
|
100–1000 μL, blue
|
100 μL
|
±3.0
|
±3.0
|
±0.6
|
±0.6
|
|
|
500 μL
|
±1.0
|
±5.0
|
±0.2
|
±1.0
|
|
|
1000 μL
|
±0.6
|
±6.0
|
±0.2
|
±2.0
|
|
0.25–2.5 mL, red
|
0.25 mL
|
±4.8
|
±0.012
|
±1.2
|
±0.003
|
|
|
1.25 mL
|
±0.8
|
±0.010
|
±0.2
|
±0.0025
|
|
|
2.5 mL
|
±0.6
|
±0.015
|
±0.2
|
±0.005
|
|
0.5–5 mL, Purple;
|
0.5 mL
|
±2.4
|
±0.012
|
±0.6
|
±0.003
|
|
|
2.5 mL
|
±1.2
|
±0.030
|
±0.25
|
±0.006
|
|
|
5.0 mL
|
±0.6
|
±0.030
|
±0.15
|
±0.0075
|
|
1–10 mL, cyan
|
1.0 mL
|
±3.0
|
±0.030
|
±0.6
|
±0.006
|
|
|
5.0 mL
|
±0.8
|
±0.040
|
±0.2
|
±0.010
|
|
|
10.0 mL
|
±0.6
|
±0.060
|
±0.15
|
±0.015
|