The G055 series instrument is an upgraded version of the 2000 version. The G055 series calculation type pH meter D-pHG2600 is used to measure the conductivity value of aqueous solutions, and then the instrument automatically calculates the pH value. This measured pH value is accurate, such as for high water quality requirements in power plants.
G055 series computational pH meterMain functions and features
1)Display the main measurement values on the same screenpHValue added conductivity and temperature;2)4.3Inch color display screen;3)Capacitive touch button;4)Chinese English menu switching, software menu friendly prompt operation;5)Output current4-20mA;6)Relay contact output;7)Communication interfaceRS485;8)Equipped with temperature automatic compensation and electrode calibration and correction functions;9)Selection of AC wide voltage and DC power supply for power supply;10)Novel appearance, good sealing, multiple installation methods, etc.
G055 series computational pH meterTwo main technical parameters
1、 Electronic unit:
1.1Measurement range and error, corresponding to conductivitypH.
Conductivity:0~20us/cmError ±1%F﹒S;
pH:7.5~10.0pHError ± 0.05;
temperature0~100℃ error ±0.5℃。
1.2Resolution:0.001,temperature0.1.
1.3Protection level:IP67.
1.4Automatic temperature compensation range:0~100℃。
1.5Isolate output current and load:4~20mA; Load<750Oh.
1.6communication interfaceRS485 Modbus RTU.
1.7Relay contacts:3A240VAC,6A28VDCor120VAC.
1.8Power supply:85~260VAC,50~60Hzor24VDC.
1.9Quality: Approximately800g.
1.10Overall dimensionsmm:145×145×138Deep, embedded (disk mounted) installation hole size:130×130contain138×138.
There are multiple installation options available: this table is compatible with wall mounted or can be customized as an all-in-one cabinet machine.
1.11Working conditions: Environmental temperature:0~45℃; Relative humidity: not greater than90%;There is no interference except for the Earth's magnetic field.
2、 Sensor electrode:
2.1Manufacturing materials for conductivity cell electrodes316L.
2.2 Electrode constant0.01.
2.3Enduring water sample pressure:0.6MPa.
2.4The installation methods of electrodes include flow type, pipeline type, and insertion type, and special installation methods can be customized.