The G055 series instrument is an upgraded version of the 2000 version. The G055 series online five parameter instrument KL5000 is widely used for continuous measurement or control of turbidity, residual chlorine, dissolved oxygen, conductivity, and pH in aqueous solutions, such as tap water, surface water, various water treatment monitoring, hospital sewage, and various wastewater discharges.
G055 series online five parameter instrumentMain functions and features
1)KL5000The instrument is equipped with five sensor electrodes of different properties, which measure five parameters;2)4.3Inch color display screen;3)Capacitive touch button;4)Chinese English menu switching, software menu friendly prompts;5)Output current4-20mA;6)Relay contact output;7)Communication interfaceRS485;8)Equipped with automatic temperature compensation and separate electrode calibration and correction functions;9)Selection of AC wide voltage and DC power supply for power supply;10)Novel appearance, good sealing, multiple installation methods, etc.
G055 series online five parameter instrumentTwo main technical parameters
1、 Electronic unit:
1.1Measurement range and error
Turbidity:0~100NTU, 0~400NTU,0~4000NTU;Error ± 5%.
Residual chlorine:0~10mg/L;Error ±0.1mg/L.
Dissolved oxygen:0~20mg/L;Error ±0.1mg/L.
Conductivity:0~200us/cm.0-2000us/cm;Error ±1%F﹒S.
pH:0.0~14.00Ph;Error ± 0.05.
temperature-10~80℃; Error ±0.5℃。
1.2Automatic temperature compensation range:0~100℃。
1.3Isolate output current and load:4~20mA; Load<750Oh.
1.4communication interfaceRS485 Modbus RTU.
1.5Relay contacts:3A240VAC,6A28VDCor120VAC.
1.6Protection level:IP67.
1.7Power supply:85~260VAC,50~60Hzor24VDC.
1.8Quality: Approximately800g.
1.9Overall dimensionsmm:145×145×138Deep, embedded (disk mounted) installation hole size:130×130contain138×138.
There are multiple installation options available: this table is compatible with wall mounted or can be customized as an all-in-one cabinet machine.
1.11Working conditions: Environmental temperature:0~45℃; Relative humidity: not greater than90%;There is no interference except for the Earth's magnetic field.
2、 Sensor electrode:
2.1)KL680ZType online turbidity sensor;2.2)CL30PType online residual chlorine electrode;2.3)028W3PType online oxygen electrode;2.4)D608PType online conductivity electrode;2.5)P026W3PType OnlinepHElectrode.