GB9706.1 standard Jiayi JAY-1167 constant temperature and humidity test chamber (desktop type)

NegotiableUpdate on 03/27
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The above prices are for reference only
Product Details

The above prices are for reference only. Interested parties please contact us by phone


1、 Product Name

Constant temperature and humidity test chamber (desktop type)

JAY-1167

Volume and size

PRODUCT MODEL

Studio Size

External dimensions

price

22L

JAY-XHW22L

25 × 30 × 30 (depth × width × height) (cm)

Approximately 110 × 55 × 105 (depth × width × height) (cm)


64L

JAY-XHW64L

40 x 40 x 40 (depth x width x height) (cm)

Approximately 133 × 62 × 115 (depth × width × height) (cm)


Note: This quotation includes shipping and on-site training fees;

Sample limitations

This experimental equipment prohibits:

Testing and storage of samples of flammable, explosive, and volatile substances

Testing and storage of corrosive substance samples

Testing or storage of biological samples

Testing and Storage of Strong Electromagnetic Emission Source Samples

performance index

Test environment conditions

The ambient temperature is 5-35 ℃ and the relative humidity is ≤ 85% RH

Atmospheric pressure and atmospheric pressure

Air quality: No high concentration dust or corrosive gases, and cannot be tested or stored in flammable and explosive gases

tap-water

Tap water requirements:

Water pipe: 1/2 "Water pressure: 0.1-0.3 Mpa

temperature range

-20℃ ~150℃

Humidity range

30%~95%RH

temperature fluctuation

±0.5℃

Humidity fluctuation

±2.5%RH

Temperature deviation

±2.0℃

Humidity deviation

≤± 2%R.h gt; 75%),≤± 2.5%R.h. lt; 75%)

heating up time

-About 65 minutes from 20 ℃ to 150 ℃

Cooling time

20 ℃~-20 ℃ for about 60 minutes

Noise

60dB

Experimental function

Constant temperature setting, constant humidity and heat, temperature rise, temperature drop, high and low temperature cycling test

Meet the testing standards

1. GB10592-89 Technical Conditions for High and Low Temperature Test Chamber

2. GB10586-89 Technical Conditions for Damp Heat Test Chamber

3. GB2423.1-2001 (IEC68-2-1) Test A: Low temperature test method.

4. GB2423.2-2001 (IEC68-2-2) Test B: High temperature test method.

GB2423.3-93 (IEC68-2-3) Test Ca: Constant Damp Heat Test Method.

8. GJBl50.9-93 (MIL-STD-810D) Wet Heat Test Method.


2、 Structural features

Structural method

Thermal insulation enclosure structure

Integrated and modular design

The equipment is divided into the following parts: enclosure, air circulation, refrigeration, heating system, humidification system, and electrical control

Outer wall material: High strength cold-rolled steel plate with double-sided fine powder baking paint or SUS304 mirror stainless steel plate

Inner wall material: SUS304 mirror stainless steel plate

Insulation material: Special composite with special insulation performance Insulation material: Hard polyurethane board and glass wool

Standard configuration of test box

1 power cord (2 meters) (can also be adjusted according to the user's on-site situation)

1 over temperature protector

1 cable hole (Φ 50) (placed on the left side of the equipment for measurement, equipped with a stainless steel cover)

There is a tempered glass door inside the double-layer outer box door, which is convenient for personnel to operate.

bearing

The load-bearing capacity of the sample rack shall not be less than 20KG/layer

Observation window

One heating wire placed on the door, which automatically opens during wet heat and low temperature tests to prevent condensation and frosting, ensuring that the equipment can observe the test situation inside the box through the observation window even when it is in the open state

light

The box door is equipped with one moisture-proof lighting lamp for easy observation of the test samples inside the box. The lighting fixtures are controlled by switches on the control panel outside the box

Control operation interface

Touch control screen, main power switch, buttons, over temperature protector, timer, etc

machine room

Refrigeration unit, power distribution control cabinet, exhaust fan, etc

Distribution control cabinet

Cooling fan

Distribution board

balance device

The box is equipped with a pressure balancing device to ensure that the pressure inside the box remains basically the same as the outside during heating, cooling, and constant testing, preventing damage to the laboratory body caused by excessive or insufficient pressure inside the box

3、 Air circulation section

fan

The power source for mixing air inside the box is the fan, which mixes the air heated by the heater inside the box and then sends it out of the channel to circulate the air inside the box

Air duct

Including air vents and return vents with air guidance, to achieve efficient heat exchange for the equipment

heater

The main heat source of the equipment provides the necessary heat for heating and maintaining a constant temperature, with a reasonable surface heat load configuration, no open flames, and a long service life

4、 Refrigeration section

The refrigeration section is the main part of the equipment that generates a cold source, providing the required cooling capacity for equipment cooling, low temperature, constant temperature, and humid heat. Depending on the testing status of the equipment, the refrigeration system automatically starts to provide cooling capacity for the corresponding testing process, thereby achieving the goal of meeting the performance indicators of the equipment

compressor

The imported air-cooled compressor is used as the power source for the refrigeration system, and the refrigeration system of this equipment adopts a cascade refrigeration system

heat exchanger

Using finned heat exchangers to provide cooling for the refrigeration system and a place for cold and hot exchange of refrigerant

evaporator

The main cold source inside the box is the main component that exchanges energy between the cooling capacity of the refrigeration unit and the hot air inside the box

throttling device

The combination of expansion valve and capillary tube is used for combined control as the throttling device of this equipment, allowing the state of the refrigerant to be changed after passing through this device

refrigerant

Using green and environmentally friendly refrigerant R404A as the refrigerant medium for this refrigeration system

Humidification section

The humidification part is the main component that generates wet steam in the equipment

It mainly consists of the following parts:

Armored heater

The water level switch

It is the main heat source for generating water vapor, which is heated according to the needs of the equipment to vaporize water and produce the steam required by the equipment

Adopting a float switch with high reliability, it provides signals for controlling the reliable addition of water to the humidifier and preventing dry burning alarms

control system

controller

Adopting a color touch imported temperature and humidity program controller, displaying in both Chinese and English

Operation mode

Fixed value/program mode: Users can conduct constant tests according to the requirements and needs of the experiment

Setting method

key-in

Input

Thermal resistance signal

output

Multi point temperature and humidity signal output

communication interface

RS232

control mode

Control and adjust the temperature and humidity inside the test chamber, automatically determine the working status of each executing component (such as compressor, fan, heater, waterway, etc.)

PID action and automatic calculation function

Ratio band (PB): 0.1% to 999.9%

Integration time (IT): SR-92/SR-93 1~3600 seconds

Differential time (DT): SR-92/SR-93 1-1200 seconds

Setting Range

Temperature: -100~+300 ℃, humidity: 0-100% RH

Control accuracy

Temperature: 0.1 ℃, Humidity: 0.1% RH, Time: 0.1 minutes

measuring sensor

Armored platinum resistance high-precision temperature sensor

timer

Equipped with an independent working time accumulation timer

5、 Other configurations


Main power leakage circuit breaker

Denard, AC220V、50Hz、 Single phase+protective earth wire

6、 Safety protection device

The equipment has multiple alarm protection functions such as overheating, fan overheating, overcurrent, compressor overpressure, and overheating. The fault self diagnosis function of the controller ensures that once the equipment malfunctions, the power supply to the main components will be cut off and an alarm signal will be issued, including:

(1) Air switch (to prevent overcurrent and short circuit);

(2) Fan overheating protection (protects the supply fan to prevent excessive fan temperature);

(3) Compressor overpressure, overheating, and overcurrent protection (protecting the compressor from excessive pressure, temperature, and current);

(4) Overtemperature protection (to prevent excessive temperature in the laboratory and protect the test sample);

(5) Local temperature protection in the air conditioning room (to prevent excessive temperature around the heating part and protect the laboratory shell);

(6) Temperature fuse dry burning protector (to prevent boiler from burning due to water shortage);

(7) Water level protection (to prevent water shortage in boilers and water supply tanks);

(8) Fuses (to prevent short circuits in control circuits and lighting circuits);

(9) Cooling water shortage alarm (to prevent cooling water shortage);

(10) Box pressure protection (to prevent damage to the laboratory body caused by excessive pressure inside the box)