KBW uses natural lighting to achieve uniform light distribution. This climate chamber ensures constant lighting and temperature in this way
(with lighting function)
Important features:
★Temperature range:0 ° C to 70 ° C
★Temperature range with illumination:5 ° C to 60 ° C
★ 3A variable positioning lighting box, each with its own5 fluorescent tubes
★ APT.line ™ Preheating chamber technology
★ Adjustable fan speed
★ Intuitive touch screen controller with time slot and real-time programming, with internal measurement value recording
★The sealed inner door is made of safety glass(ESG) made
★Through special meansTIMELESS coating prevents corrosion of glass
★Made of stainless steel3 plug-in racks
★Access hole with silicone plug30 mm, left
★Independent adjustable temperature safety device3.1 level (DIN 12880), equipped with optical and acoustic temperature alarms
★ Computer interface: Ethernet
★ Door heating
Technical Specifications:
|
model
|
The dimensions of the casing, including the installation and connection components
Width x Height x Depth
[mm]
|
internal dimensions
Width x Height x Depth
[mm]
|
Net weight
[kg]
|
Internal volume
[L]
|
|
KBW 720
|
1250 x 1925 x 890
|
970 x 1250 x 576
|
377
|
698
|