E-mail
33400586@qq.com
-
Phone
13803929335
-
Address
No. 183 Tonghua Lane, Qibin District, Hebi City
Hebi Hengfa Instrument Co., Ltd
33400586@qq.com
13803929335
No. 183 Tonghua Lane, Qibin District, Hebi City

HF-2000Coke reactivity and post reaction strength tester
HF-2000 Coke Reactivity and Post Reaction Strength TesterIt is an instrument for measuring the reactivity index of coke.This instrument is controlled by PLC,Industrial touch screen display,The temperature of the high-temperature furnace is controlled by temperature closed-loop fuzzy logic PID, and the PLC system monitors and processes the central temperature of the furnace on site. The application parameters of the PLC can be operated and modified through the touch screen, and the variable curve can be dynamically displayed.
1HF-2000Coke reactivity and post reaction strength testerofWorking environment and installation conditions
The electrical control system adopts TN-S wiring method.
The casing of electrical equipment must be reliably grounded.
Electrical equipment must be installed in an environment free from vibration, dust, and dust.
Working voltage: AC220V ± 10%, 50Hz ± 5%, three-phase five wire system.
Altitude ≤ 2000m.
Working environment: 5 ℃ ≤ ambient temperature ≤ 40 ℃, humidity ≤ 85%, no strong magnetic field, no severe vibration.
IIHF-2000Coke reactivity and post reaction strength testerofTechnical performance and parameters
Dynamically display the operating status of each part of the system.
●Industrial touch screen display
Dynamically display the sidewall temperature value of the high-temperature heating furnace.
Dynamically display the center temperature curve of the high-temperature heating furnace.
Dynamically display the gas flow curve of the gas path.
Dynamically display the temperature setting value of the high-temperature heating furnace.
Dynamically display the power output value of the high-temperature heating furnace.
Dynamically display the voltage output value of the high-temperature heating furnace.
Dynamically display the current control value of the high-temperature heating furnace.
Dynamically display the heating rate of the high-temperature heating furnace.
● It has protection function for rapid and slow heating speed.
● Equipped with temperature protection function for ultra-high temperatures.
● It has the function of interruption fault protection.
● Equipped with thermocouple fault protection function.
High temperature electric furnace: 7KW
Temperature control accuracy: 1100 ℃± 5 ℃
Temperature control range: 0-1150 ℃, resolution 1 ℃.
Temperature measuring element: N-type thermocouple
● Operating temperature: ≤ 1150 ℃
Heating element: silicon carbon rod
Gas flow accuracy: ± 1.5 ℃% F.S.
Linearity of gas flow rate: ± (0.5-1.5)% F.S.
Gas flow rate repeatability accuracy: ± 0.2% F.S.
Gas flow response time: 1 to 4 seconds.
Working voltage: AC220V ± 10%, 50Hz ± 5% three-phase five wire system.
Working environment: 5 ℃ ≤ ambient temperature ≤ 40 ℃, humidity ≤ 85%, no strong magnetic field, no severe vibration.