HFHXTX-6 Activated Carbon Denitrification Value Tester
Composition of the device:
HFHXTX-6 Activated Carbon Denitrification Value TesterIt consists of gas source, pressure reducing device, flow control system, preheater system, heating furnace system, measuring tube system, gas detection system, etc.
Main technical parameters:
Preheating furnace: 0-300 ℃, inner diameter 108mm, constant temperature zone greater than 600
Heating furnace: 200 ℃, constant temperature zone greater than 1000mm
Reaction determination tube: inner diameter 108mm
Total gas flow rate: 52.0L/min
Ammonia volume fraction: 0.02%, carbon monoxide volume fraction: 0.02%
Oxygen volume fraction: 6.4%, water inflation volume fraction: 8.0%
Out of stock temperature: 120 ± 5 ℃
Measurement principle:
A mixed gas containing nitrogen oxides, ammonia gas, nitrogen gas, and water vapor is passed through an activated coke bed, and the nitrogen oxides in the gas are selectively catalytically reduced (SCR) by ammonia gas to produce nitrogen gas emissions. When the SCR reaction reaches a steady state, the ratio of the volume fraction of bulk nitrogen oxides in the converted nitrogen oxides to the volume fraction of nitrogen oxides introduced into the gas is the denitrification rate of the activated coke.
Operating procedure:
1. Load the sample into the measuring tube and perform airtightness testing.
2. Raise the preheating furnace to 300 ℃ and the reactor to 120 ℃.
3. Adjust the flow rate of each gas, mix and introduce it into the reactor, and start timing.
4. Measure the volume fraction of NO (dry) every 120 minutes, and the difference in volume fraction of NO in the rated exhaust gas for four consecutive times should not exceed 5 * 10-6Stop measuring at that time.
5. Calculate the out of stock rate
Maintenance and precautions:
1. Due to the harmful nature of NO and ammonia gases, the device should be tested under good ventilation conditions.
2. Before measurement, the system must undergo an air tightness check, and can only be measured after the air tightness is qualified.
3. When operating the heating furnace, wear protective gloves and be careful of burns.
4. After completing the measurement, check the device and measurementClean up the premises to protect environmental hygiene
Learn moreReactivity of cokeProducts please go to/