Welcome Customer !

Membership

Help

Huai'an Pruda Testing Equipment Co., Ltd
Custom manufacturer

Main Products:

ybzhan>Products

Huai'an drug stability test box

NegotiableUpdate on 03/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Drug stability test box, mainly used as an instrument for measuring the required long-term stable temperature, humidity, and lighting environment
Product Details

Drug stability test box, main purpose of the test box

An instrument used for measuring temperature, humidity, and lighting environments that require long-term stability. Suitable for pharmaceutical companies to conduct accelerated testing, long-term testing, high-temperature testing, and strong light irradiation testing on drugs and new drugs. It is a drug stability testing selection scheme for pharmaceutical companies to conduct drug stability testing. The drug stability test box is used for intensive research in industries such as pharmaceuticals, medicine, biotechnology, food industry, electronics industry, and all related industries including life sciences. The WHO principle requires long-term stability test conditions at 25 ℃/60% RH humidity. The standard for stability testing systems in the pharmaceutical industry is to test the humidity of 40 ℃/75% RH for 6 months in accelerated testing.

Characteristics of box structure:

·The material used in the drug stability test chamber is SUS304B board, which has the characteristics of acid resistance, corrosion resistance, and easy cleaning;
·The position of the sample rack can be adjusted up and down as needed;
·The use of high-quality door seals and insulation materials enhances the overall performance of the machine;
·A reasonable air duct circulation system ensures high uniformity of temperature and humidity inside the box;
·The water supply system of the drug stability test chamber adopts automatic water replenishment function.

controller:

Both temperature and humidity are controlled by imported controllers, with precise display settings for relative temperature and humidity performance, and a resolution of 0.1 ℃/1% RH.

Compliant with standards:

The drug stability test chamber is manufactured in accordance with the 2010 Pharmacopoeia Guidelines for Drug Stability Testing and relevant provisions of GB/T10586-2006.

name Drug testing chamber Comprehensive drug testing chamber
model YP-150SD YP-150SDP YP-150GSP YP-150GSD
YP-250SD YP-250SDP YP-250GSP YP-250GSD
YP-500SD YP-500SDP YP-500GSP YP-500GSD
YP-1000SD YP-1000SDP YP-1000GSP YP-1000GSD
YP-2000SD YP-2000SDP YP-2000GSP YP-2000GSD
Capacity/inner/outer bladder size 150L / 600*405*620mm / 800*910*1630mm
250L / 500*600*830mm / 880*1010*1800mm
500L / 800*700*900mm / 1000*1210*1900mm
1000L / 1000*1000*1000mm / 1200*1400*1820mm
2000L / 2050*750*1350mm / 2210*1150*1980mm
performance index Temperature Control Range 10~65˚C
Temperature control fluctuation ±0.5˚C ±1.0℃
Temperature deviation ±1.5˚C ±1.5˚C
Moisture control range 30~95%RH 30~95%RH
Control humidity fluctuations ±2%RH ±2%RH
Humidity deviation ±3%RH ±3%RH
Light intensity Adjustable from 0 to 6000LUX
Illuminance error ≤±500LUX
Temperature and humidity regulation method BTHC
refrigeration system Cooling method Direct cooling high-efficiency refrigeration
refrigeration compressor Imported fully enclosed compressor (Danfoss)
cooler Air-cooled cooler
control system controller SD/GD Microcomputer Controller (Digital Display Instrument) SDP/GSP Programmable Controller (Touch Screen Instrument)
temperature sensor Imported temperature and humidity sensor
Ambient Temperature +5~30˚C
power supply AC 220V 50Hz
Cargo tray (standard) 2 pieces
Safety device Compressor overload protection, water shortage protection, over temperature protection, electrical short circuit protection
Main technical parameters of two box and three box (selected according to customer's actual requirements)
name Two box drug stability box Three box drug stability box
model YP-GSD (two boxes) YP-TSD (three boxes)
Temperature Control Range 10˚C~65˚C
fluctuation ≤±0.5˚C
Moisture control range 30~95%RH
fluctuation ±2%RH
Light intensity Adjustable from 0 to 6000LX
Illuminance error Upper box lighting 400 * 400 * 500 Upper box damp heat 400 * 500 * 500 Right box lighting 300 * 400 * 500
Inner container size Lower box damp heat 400 * 500 * 500 Lower box damp heat 400 * 500 * 500
refrigeration compressor Imported compressor (Danfoss)
Temperature and humidity controller GSD microcomputer controller (digital display instrument) TSD programmable controller (touch screen instrument)
humidity sensor Capacitive type