-
E-mail
76942054 @qq.com
-
Phone
18028937131
-
Address
No.15 Shihuan Road, Sangyuan Community, Dongcheng Street, Dongguan City
Guangdong Kebao Experimental Equipment Co., Ltd
76942054 @qq.com
18028937131
No.15 Shihuan Road, Sangyuan Community, Dongcheng Street, Dongguan City
Imported high and low temperature impact test chamber
Product Introduction:
Technical Specifications | |
Temperature range of the test tank |
-55℃— +100℃ |
High temperature tank temperature setting range |
+60℃— +150℃ |
Heating time of high temperature tank |
RT(Room temperature)--+150Approximately required at ℃40min(Room temperature at+)10--+30At ℃) |
Low temperature tank temperature setting range |
-65℃---10℃ |
Cooling time of low-temperature tank |
RT(Room temperature)-- -65Approximately required at ℃85min(Room temperature at+)10--+30At ℃) |
temperature fluctuation |
±1.0℃ |
Temperature Uniformity |
±2.0℃ |
Impact recovery time |
-55--+85Approximately required at ℃5min.※ The constant temperature time for high and low temperature impact is30minAbove ※ |
product structure | |
Inner box dimensions |
W×H×D :500×500×400(mm) |
outer box dimensions |
W×H×D :1800×2000×2200(mm) |
Inner box material |
Mirror stainless steel plate(SUS#304) |
outer box material |
Sandgrain stainless steel plate(SUS#304)Or steel plate baking paint |
Insulation material |
a.High temperature tank: Aluminum silicate insulation cottonb.低温槽:高密度PUfoaming |
door of the box |
Single panel doors for upper, middle, and lower boxes,left open. a.Embedded flat handleb.Rear button:SUS#304 c.Silicone foam strip |
Sample placement rack |
a.Intermediate testing roomb.Load bearing capacity not exceeding5KG c.stainless steelSUS#304inner box |
heating system |
Fin type heat dissipation tube shaped stainless steel electric heater 1.High temperature tank6 KW 2.Low temperature tank3.5 KW |
Air circulation system |
1.motor1HP×2Taiwan 2.Stainless steel extended axis 3.Multi wing fan blade(SIROCCO FAN) 4.Specially designed fan forced air circulation system |
Imported high and low temperature impact test chamberFreezing system | |
compressor |
French "Taikang" brand fully enclosed compressor |
condenser |
Fin type attached platform dispersing hot air fan |
evaporator |
Fin type multi-stage automatic load capacity adjustment |
expansion valve |
Danfoss components from Denmark |
refrigerant |
US imports new environmentally friendly refrigerantsR404A |
Refrigeration auxiliary components |
Desiccant (Denmark),Oil separator (USA),Refrigerant windows (Denmark), solenoid valves (Japan), filters (Japan), pressure controllers (South Korea), shut-off valves (Italy) and other refrigeration accessories are all imported parts |
Function:
Used in industries such as electronic and electrical components, automation parts, communication components, automotive parts, metals, chemical materials, plastics, etc., as well as in the defense industry, aerospace, and military industryBGAThePCBBase plate, electronic chipICThe physical changes of semiconductor ceramic magnets and polymer materials, testing their repeated resistance to high and low temperatures, as well as the chemical changes or physical damage caused by thermal expansion and contraction of the product, can confirm the quality of the product from a precise perspectiveICIt is used in components of heavy machinery and is a testing box for product testing in various fields.
Features:
1Simple control panel, full-color touch screen control, temperature and humidity can be programmed simultaneously.
2Adopting the original French "Taikang" compressor circulation system, the operation is more stable and reliable.
3Glass observation window, which can be used for observing test samples in experiments, equipped with anti sweat electric heating device to prevent water vapor condensation and droplets, and high brightness fluorescent lamp to maintain illumination inside the box.
4Stainless steel inner box, easy to operate, low noise.

composition
It refers to a test chamber that can simultaneously apply temperature and humidity stress. Composed of refrigeration system, heating system, control system, temperature and sensor system, etc.
Satisfied test methods:
GB/T2423.1.2TheGB/T10592-2008TheGJB150.3High and low temperature impact test chamber.
Precautions for high and low temperature impact test chamber:
1Keep away from heat sources such as stoves and radiators. Meanwhile, direct sunlight should be avoided. This is beneficial for high and low temperature impact test chambers.
2It should be placed in a stable location on the ground. Otherwise, the compressor will generate vibration and a large amount of noise during startup, which will shorten the service life of the high and low temperature impact test chamber for a long time.
3It should be placed in a well ventilated area. The distance between the back of the test box and the wall should be greater than60Centimeters, the top height should be greater than60Centimeters. There should not be too much debris around.
Guangdong Kebao Testing Equipment Co., Ltd. has been focusing on constant temperature and humidity test chambers, constant temperature and humidity machines for many yearsthermal shock chamber、High and low temperature impact test chamberEquipment such as salt spray test chamber, ultraviolet aging test chamber, and sand and dust test chamber. Welcome to call for consultation: