E-mail
dan600507@163.com
-
Phone
19916557244
-
Address
Room 1706, International Trade Building, No. 55 Keelung Road, Waigaoqiao Free Trade Zone, Shanghai
Shanghai Lianjun Engineering Equipment Co., Ltd
dan600507@163.com
19916557244
Room 1706, International Trade Building, No. 55 Keelung Road, Waigaoqiao Free Trade Zone, Shanghai
Labom type bimetallic thermometer
Bimetallic thermometer,Equipped with an adjustable connector system, model series FA
Design features
——The material of the fusion part between the shell and the medium is stainless steel
——Dial DN100/160, protection level IP66
——Can provide different forms of connections
——Accuracy level 1 meets EN 13190 standard
——Fine tuning the pointer for correcting the indication
——Indicator damping (standard)
optional
——Dial filling
——Material certification in accordance with DIN EN 10204
——Can be connected to Zone 0 with protective sleeve
Application
These thermometers are suitable for outdoor and corrosive environments. Other liquid damping can also be provided for extreme temperature conditions. The temperature probe is prone to bending, so we recommend installing a suitable protective tube.
Application Fields
——Chemistry and Petrochemicalsindustry
——Mechanical processing
——Main process technology
——Ship
Technical data
housing
Stainless steel No. 1.4301(304),
Dial sizes 100 and 160 mm
Labom type bimetallic thermometer
Shell design
Protection level IP66 in accordance with EN60529
Can be filled with liquid at the same time
Measuring element
The constant temperature bimetallic spiral conforms to DIN 1715 and has good adjustment force, fast action, and laser welding for heat aging, substrate, and connectors
Pointer axis
Stainless steel No. 1.4571(316Ti)Equipped with numerous bearings and flexible joint screws
Temperature measurement kit
Stainless steel No. 1.4571(316Ti)Diameter 8 or 6 mm
Can provide standard length,Refer to the ordering instructions
Front glass
Instrument glass or polymer materials
Shell sealing
DINNA rubber
Measurement system damping
As a standard, the indicator damping can provide additional damping filling for the entire system, with scales ranging from -40 ℃ to 160 ℃
Nominal range
Complies with NE13190 from -40...+500 ℃ (with a limit of 600 ℃) and other values as required. For ranges above 400 ℃, we recommend using protective tubes that comply with DIN 43772
precision
Compliant with EN13190 Level 1. These levels are not caused by errors, but by changing the position of the connection point. However, this possible error can be compensated for by readjusting the adjustable pointer
Storage or transportation temperature
Z-20...+60 ℃, other values as required
Application Guide
The load capacity of the temperature measurement package depends on the following parameters:
If required, necessary technical testing can be conducted