The G055 series instrument is an upgraded version of the 2000 version. The KL2266 online conductivity+pH multi parameter instrument is suitable for power plants and is used for online continuous measurement and analysis of pH and conductivity values of various aqueous solutions, as well as automatic acid-base control. It is also used for surface water, tap water, food and beverage pharmaceuticals, chemical, metallurgical, mining, reverse osmosis, industrial process measurement and dosing control, various water treatment and environmental monitoring; It is particularly suitable for measuring the pH value and conductivity value of condensed water from power plant generators, and for measuring high-purity water, purified water, ordinary water, and wastewater over the full range.
Online conductivity+pH multi parameter meterDetailed introduction:
Main functions and features
1)2-in-1: One instrument with two electrodes for measuring and displaying conductivity andpHAnd temperature; Other parameters can also be added and combined into two parameters, three parameters, four parameters, and five parameters;2)4.3Inch color display screen;3)Capacitive touch button;4)Chinese English menu switching, software menu friendly prompt operation;5)Recorder function, long-term storage of measurement data;6)Independent output current4-20mA;7)Independent relay contact output;8)Communication interfaceRS485;9)Equipped with automatic temperature compensation and multiple electrode calibration functions;10)Selection of AC wide voltage and DC power supply for power supply;11)Good sealing performance, multiple installation methods, etc.
Online conductivity+pH multi parameter meter
Two main technical parameters
1、 Electronic unit:
1.1Measurement range and error
Conductivity:0.055~2~20us/cmError ±1% F·S;0.055~200us/cmError ±1%F·S;0.055~2ms/cmError ±3%F·S; 0.055~100ms/cmError ±3%F·S.
pH:0.0~14.00pHError ± 0.01;compatibleORPMeasurement ±1999mV±5mV.
temperature-20~130℃ error ±0.5℃。
1.2Resolution:0.001.
1.3Protection level:IP67.
1.4Automatic temperature compensation range:0~100℃。
1.5Isolate output current and load:4~20mA(Independent two-way output measurement of main value and temperature); load<750Oh.
1.6communication interfaceRS485 Modbus RTU.
1.7Relay contacts:3A240VAC,6A28VDCor120VAC(Three groups).
1.8Power supply:85~260VAC,50~60Hzor24VDC.
1.9Quality: Approximately800g.
1.10Overall dimensionsmm:145×145×138Deep, embedded (disk mounted) installation hole size:130×130contain138×138.
There are multiple installation options available: this table is compatible with wall mounted or can be customized as an all-in-one cabinet machine.
1.11Working conditions: Environmental temperature:0~45℃; Relative humidity: not greater than90%;There is no interference except for the Earth's magnetic field.
2、 Sensor electrode:
Sensor electrodes are configured according to the requirements of the on-site medium, refer toDDG2022Industrial conductivity meter andpHG2026Type IndustrypHExplanation within the data.