PT100/PT1000/thermocouple six channel temperature control instrument (can be equipped with RS485)
1、 Technical performance
1. Input types: CU50 (-50.0-150.0 ℃), Pt100 (-199-600.0 ℃), Pt1000 (-199-600.0 ℃)
K(0~1300℃)、E(0~700.0℃)、J(0~900.0℃)、S(0~1600℃)
Note 1: When the user places an order specifying input signals of 0-5V and 1-5V analog signals, the instrument no longer supports the input of various thermocouple and thermistor signals mentioned above,
Note 2: This instrument is a multi-channel instrument that can connect multiple sensors. But isolated sensors must be selected. Taking a thermocouple as an example: both wires of the thermocouple
Cannot conduct with metal casing. Otherwise, there may be situations such as inaccurate temperature measurement.
2. Control methods: Two position control, PID control
3. Output mode: Relay 240V 5A (resistive load) SSR trigger signal 20ma/9V
4. Measurement accuracy: ± 0.5F • S ± 1 word, additional cold end calibration error ± 1 ℃
5. Working power supply: AC220V 50Hz Power consumption: less than 5W
6. Working environment: 0~50 ℃, relative humidity ≤ 85% RH, non corrosive and non strong electromagnetic radiation environment
7. Dimensions: 160 × 80 × 120 Hole size: 152 × 76 (mm)
2、 Instrument panel
3、 Instrument wiring (for reference only, based on the wiring diagram on the instrument)
Note: When communicating via RS485, the instrument should be connected to 'VA+' and 'VB -';
4、 Instrument usage
1. First level menu settings
Press the function key (SET key) for 3 seconds to enter the first level menu. At this time, the 'First Display Window' and 'Second Display Window' will display parameter symbols and values respectively. You can press the ◄ (cursor shift key), ▲ (plus key), and ▼ (minus key) keys to change the parameter values. After the modification is complete, press the SET key to save and enter the next parameter; Modify other parameters using the same method.
2. Secondary menu settings
Press CH1, CH2, CH3, CH4, CH5, and CH6 for three seconds to enter the corresponding channel menu items for each channel parameter. Then press the ◄, ▲, and ▼ keys to change the parameter values. After completing the modification, press the SET key to save and enter the next parameter; The parameters are shown in the table below:
|
First level menu, press the function key (SET key) for 3 seconds to enter |
|||||||||
|
serial number |
prompt |
Storage address |
name |
set the scope |
Instructions |
factory default |
|||
|
one level dish single |
0 |
LOCK |
0000H |
combination lock |
0~50 |
When the password lock is set to 18, all modifications are allowed Parameters are prohibited from being modified unless they are set to 18 |
18 |
||
|
1 |
Sn |
0001H |
Enter specifications |
0~5 |
CU50ThePtTheKTheETheJTheS |
K |
|||
|
2 |
t |
0002H |
Output cycle |
0~120 S |
Set the action cycle for relay control |
20 S |
|||
|
3 |
dp |
0003H |
decimal point position |
0~1 |
0. No decimal point; 1. There is a decimal point |
0 |
|||
|
4 |
DIL |
0004H |
Lower limit of set value |
Range starting point~DIH |
This parameter limits the lower limit of the set value |
random |
|||
|
5 |
DIH |
0005H |
Set the upper limit of the value |
DIL~Full range |
This parameter limits the upper limit of the set value |
random |
|||
|
6 |
OPB |
0006H |
Secondary output mode |
0~2 |
0.noneoutput 1.openRS485 serial communicationfunction |
0 |
|||
|
7 |
Add |
0007H |
communicationaddress |
1~64 |
The numbering of instruments in the centralized control system |
1 |
|||
|
8 |
bt |
0008H |
communicationBaud rate |
0~3 |
0:1200; 1:2400; 2:4800;3:9600 |
9600 |
|||
|
Secondary menu, press CH1 key for 3 seconds to enter 1Channel menu |
|||||||||
|
two level menu
one Channel parameter setting |
9 |
SP1 |
0009H |
1-channel control point setting |
Scope byDILTheDIHdecide |
Temperature setting parameters for channel 1 control point |
random |
||
|
10 |
AL1 |
000AH |
Alarm setting value |
-- |
By ALP1Parameters determine the alarm mode For details, seeALP1Parameter Description |
0 |
|||
|
11 |
SC1 |
000BH |
Error correction value |
±20.0 |
Sensor errortranslationcorrection value |
0 |
|||
|
12 |
P1 |
000CH
|
Proportional band |
0~200.0 |
adoptPIDWhen controlling,The proportional band determines the magnitude of the system's proportional gain, and the larger P, The smaller the effect of proportion, the smaller the overshoot, but too small will increase the heating time. (When P=0,The instrument willConvert to two position control,Some actuators that are not suitable for frequent start-up, such as fans, compressors, motors, etc., should use this method. Please refer to the parameters for detailsHY1The explanation section) |
15.0 |
|||
|
13 |
I1 |
000DH |
integration time |
0~3000 |
Set the integration time to release proportional control The residual deviation that occurs will delay the time for the system to reach equilibrium if it is too large,Too small can cause fluctuations |
240 |
|||
|
14 |
D1 |
000EH |
Differential time |
0~200S |
Set differential time to prevent output waves Move to improve the stability of control |
30 |
|||
|
15 |
Hy1 |
000FH |
Main control and alarm return difference |
0.1~50.0 |
When the instrument is a two position control(Proportional tape)P=0)The dead zone between the opening or closing of the actuator to avoid frequent actions of the actuator. When the main control is in heating mode: whenmeasured valuePV1≤set valueSV1-HY1When the actuator opens, whenPV1≥SV1+HY1When, the actuator closes. |
1.0 |
|||
|
16 |
At1 |
0010H |
Self-tuning parameters |
0~1 |
0: Disable self-tuning 1: Enable self-tuning |
0 |
|||
|
17 |
CoL1 |
0011H |
Heating and coolingmethodchoice |
0~1 |
0: Heating function 1: Cooling function |
0 |
|||
|
18 |
ALP1 |
0012H |
Alarm definition |
0~4 |
0: No alarm; 1: Upper limit alarm 2: Lower limit alarm;3:Upper deviation alarm 4Deviation alarm 5Alarm outside the interval 6Alarm reported within the interval |
1 |
|||
|
Press separatelyCH2/CH3/CH4/CH5/CH6Press the key to enter2-6Channel menu,Set up using the same method |
|||||||||
|
Press CH6Press the key for 3 seconds to enter 6Channel menu |
|||||||||
|
sixChannel parameter setting |
59 |
SP6 |
003BH |
Ditto. |
|
|
|
||
|
60 |
AL6 |
003CH |
Ditto. |
|
|
|
|||
|
61 |
SC6 |
003DH |
Ditto. |
|
|
|
|||
|
62 |
P6 |
003EH |
Ditto. |
|
|
|
|||
|
63 |
I6 |
003FH |
Ditto. |
|
|
|
|||
|
64 |
D6 |
0040H |
Ditto. |
|
|
|
|||
|
65 |
Hy6 |
0041H |
Ditto. |
|
|
|
|||
|
66 |
At6 |
0042H |
Ditto. |
|
|
|
|||
|
67 |
CoL6 |
0043H |
Ditto. |
|
|
|
|||
|
68 |
ALP6 |
0044H |
Ditto. |
|
|
|
|||
Note: Please refer to the "Communication Protocol" section later for the measurement value addresses of each channel.
5、 Self tuning operation
First, set the settings and then enter the menu,Set the hysteresis to around 0.5-1, set AT to ON, and the instrument enters a self-tuning state. The corresponding display window flashes and displays the "AT" character. At this time, the instrument is in positional control. After three oscillations, the instrument determines new parameters such as P, I, D and saves them, and resets to the control state.
sixError message
1.
When the instrument window displays "HH or LL", please check if the sensor wiring is correct.
2.