E-mail
13901452326@139.com
-
Phone
17798959858
-
Address
Chuangye Road, Baoying Liubao Industrial Concentration Zone, Yangzhou City, Jiangsu Province
Yangzhou Zhongyue Electric Co., Ltd
13901452326@139.com
17798959858
Chuangye Road, Baoying Liubao Industrial Concentration Zone, Yangzhou City, Jiangsu Province
Parameters of intelligent dielectric loss tester
Product Introduction:
The dielectric loss tester is used for on-site anti-interference dielectric loss measurement or laboratory precision dielectric loss measurement. The instrument is an integrated structure with built-in dielectric bridge, variable frequency power supply, test transformer, and standard capacitor. Adopting frequency conversion anti-interference and Fourier transform digital filtering technology, the measurement is fully automated and intelligent, and the measurement data is very stable under strong interference. The measurement results are displayed on a large LCD screen and come with a built-in micro printer for printing and output.
Working Principle:
The instrument measurement circuit includes a standard circuit and a test circuit, as shown in Figure 3.
The standard circuit consists of a built-in high-voltage stability standard capacitor and a standard resistor network. The computer collects the current amplitude and phase difference of the standard circuit and the test circuit in real time, and calculates the capacitance value (CX) and dielectric loss (tg δ) of the tested product.
All data acquisition circuits use high-voltage stabilizing devices, and the acquisition board and acquisition computer are shielded by an iron box floating in the air, while the instrument casing is grounded and shielded; In addition, anti-interference technologies such as optical data, floating ground, large-area ground, single point ground, and digital filtering were used. In addition, the computer processed data from hundreds of power grid cycles, making the measurement results stable, accurate, and reliable.
Technical Indicators:
1. Measurement of dielectric loss and capacitance
Accuracy: Cx: ± (reading x 1%+1pF)
tgδ: ± (reading x 1%+0.00040)
Anti interference indicators: Variable frequency anti-interferenceCan still achieve the above accuracy under 200% interference
Range of capacitance: Apply high pressure internally: 3pF~60000pF/10kV 60pF~1.2μF/0.5kV
External high voltage: 3pF~1.5 μ F/10kV 60pF~30 μ F/0.5kV
resolution: 0.001pF,4 significant digits
Range of TG δ: Unlimited, resolution0.001%, automatic identification of three types of test samples: capacitor, inductor, and resistor.
Test current range: 10 μ A~5A
Apply high pressure internally: Set voltage range: 0.5~10kV
Output current: 200mA
Boosting and lowering method: Continuous smooth adjustment
Voltage accuracy: ± (1.5% x reading+10V)
Voltage resolution: 1V
Test frequency: 45~65HzInteger frequency
49/51Hz、45/55HzAutomatic dual frequency conversion
Frequency accuracy: ±0.01Hz
External high pressure: Test current 5A during positive wiring/40~70Hz
Test current 10kV during reverse wiring / 5A /40~70Hz
CVT self-excited low-voltage output: output voltage 3~50V, output current 3~30A
Measurement time: about30s,Related to the measurement method
2. Other indicators
Input power supply: 180V~270VAC,50Hz/60Hz±1%Powered by mains electricity or generators
Computer interface:Standard RS232 interface
printer:bring your ownMicro thermal printer
Ambient temperature -10℃~50℃
Relative humidity: <90%,No condensation
Parameters of intelligent dielectric loss tester