Phone
13761574431
-
Address
No. 229 Jinyuan Fourth Road, Jiading District, Shanghai
Shanghai Xiwei Scientific Instrument Co., Ltd
13761574431
No. 229 Jinyuan Fourth Road, Jiading District, Shanghai
Temperature range: ambient temperature plus5 ° C to 100 ° C
utilizeAPT.line ™ Technology achieves high temperature accuracy
Free convection
beltController for LCD display
Exhaust valve electromechanical control
Made of safety glassESG made inner door
2 chrome plated sockets
Integrated independent adjustable temperature safety device3.1 level (DIN 12880), using optical alarm
Ergonomic handle design
Used for recording dataUSB interface
| name | BD720-230V¹ | BD720UL-240V¹ |
| Choose a model | standard | standard |
| Item code | 9010-0331 | 9010-0332 |
| measure | ||
| Internal volume [L] | 737 | 737 |
| Net weight of equipment (unloaded) [kg] | 169 | 169 |
| Total load [kg] | 315 | 315 |
| Load per partition [kg] | 45 | 45 |
| The dimensions of the casing do not include additional components and connectors | ||
| Net weight width [mm] | 1165 | 1165 |
| Net weight at height [mm] | 1590 | 1590 |
| Depth net weight [mm] | 870 | 870 |
| Wall to side distance [mm] | 100 | 100 |
| Wall distance from rear [mm] | 160 | 160 |
| Internal dimensions | ||
| Width [mm] | 960 | 960 |
| Height [mm] | 1280 | 1280 |
| Depth [mm] | 600 | 600 |
| The number of doors | ||
| Inner door | 2 | 2 |
| Outer door | 2 | 2 |
| Data related to the environment | ||
| Energy consumption at 37 ° C [Wh/h] | 78 | 78 |
| firmware | ||
| Number of partitions (hours/) | 2/16 | 2/16 |
| Temperature performance data | ||
| The temperature range exceeds the indoor temperature by 5 ° C to [° C] | 100 | 100 |
| The temperature deviation is 37 ° C [± K] | 0.7 | 0.7 |
| Temperature fluctuation at 37 ° C [± K] | 0.1 | 0.1 |
| Heating time to 37 ° C [min] | 70 | 70 |
| The recovery time after opening the door in 30 seconds is 37 ° C [min] | 23 | 23 |
| Electric data | ||
| Rated voltage [V] | 230 | 240 |
| Power frequency [Hz] | 50/60 | 50/60 |
| Rated power [kW] | 1.65 | 1.75 |
| Equipment safety device [A] | 16 | 16 |
| Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable toStandard specification unloaded equipment under ambient temperature of 22 ± 3 ° C and ± 10% power supply voltage fluctuation. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. We reserve the right to make technical changes.