The portable multifunctional calibrator MC2-MF (pressure calibrator, temperature calibrator, signal calibrator) is a highly practical portable on-site thermal calibration tool. It is a product of Finland's Beamex company and has all the functions of temperature, pressure, and process signal measurement and calibration: it has a built-in pressure module and can choose multiple external pressure modules to expand the pressure range; Measurement and simulation of resistance, thermistor RTD, thermocouple TC, voltage, millivolt mV, current, pulse, and frequency, switch testing, HART compatibility, supplying 24V loop power. Chinese menu operation interface. Large screen, dual windows.
MC2-MF Portable multifunctional calibrator
(BEAMEX MC2-TE)
|
Multi functional for measuring and verifying temperature, pressure, and electrical signals:high-precision
u Built in pressure module
u Multiple external pressure modules can be selected to expand the pressure calibration range
u thermal resistanceRTDMeasurement and simulation generate
u thermocoupleTCMeasurement and simulation generate
u Resistance measurement and simulation generation
u millivoltmVMeasurement and simulation generate
u Voltage generation
u Current generation (internal and external power supply)
u Pulse generation
u Frequency generation
u Current measurement (internal and external power supply)
u voltage measurement
u frequency
u PULSE COUNT
u internalHARTcompatible
u supply24VLoop Power
u Switch test
u Chinese menu operation interface
|
|
MC2-MFTechnical parameter indicators —— MC2-MFPortable multifunctional calibrator
|
characteristic
|
describe
|
|
display
|
60mm×60mm;LCDdisplay
|
|
weight
|
770g
|
|
size
|
215mm×102mm×49mm(d/w/h)
|
|
keyboard
|
membrane keyboard
|
|
Type
|
rechargeableNiMH,4000mAh,3.6V DC
|
|
charging time
|
5hour
|
|
Charging power supply
|
100~240VAC,50~60Hz
|
|
Battery settings
|
All day measurement mode, background light off
Half day mode provided12mACircuit power supply, background light on
|
|
Optional dry battery settings
|
8Hour measurement mode, background light off
4Hour measurement mode provided12mALoop Power,Background light on
|
|
operation temperature
|
-10~50℃(14~122℉)
|
|
storage temperature
|
-20~60℃(-4~140℉)
|
|
humidity
|
0arrive80%non-condensing
|
|
Max input voltage
|
30V DC,60V DC
|
|
safety standards
|
Derective 73/23/EEC,EN61010-1
|
|
EMC
|
Derective 89/336/EEC,EN61010-1
|
1Measurement and verification of electrical signals —— MC2-MFPortable multifunctional calibrator
voltage measurement(-1~60V DC)
|
range
|
resolution
|
1Annual uncertainty
|
|
±0.25V
|
0.001mV
|
0.02%RDG+5uV
|
|
±(0.25~1V)
|
0.01mV
|
0.02%RDG+5uV
|
|
1~25V
|
0.1mV
|
0.02%RDG+0.25mV
|
|
25~60V
|
1mV
|
0.02%RDG+0.25mV
|
Milliampere measurement(100mA)
|
range
|
resolution
|
1Annual uncertainty
|
|
±25mA
|
0.0001mA
|
0.02%RDG+1.5uA
|
|
±(25~100mA)
|
0.001mA
|
0.02%RDG+1.5uA
|
Loop Power
|
characteristic
|
parameter
|
|
output current
|
MAX25mA,There is short circuit protection
|
|
Output voltage
|
24±10%
|
|
HARTOutput impedance in compatibility mode
|
300ohm±20%
|
frequency(0.0027~50,000Hz)
|
range
|
resolution
|
1Annual uncertainty
|
|
0.0027~0.5Hz
|
0.000001 Hz
|
0.01%RDG
|
|
0.5~5 Hz
|
0.00001 Hz
|
0.01%RDG
|
|
5~50 Hz
|
0.0001 Hz
|
0.01%RDG
|
|
50~500 Hz
|
0.001 Hz
|
0.01%RDG
|
|
500~5000 Hz
|
0.01 Hz
|
0.01%RDG
|
|
5000~50000 Hz
|
0.1 Hz
|
0.01%RDG
|
Pulse counting
|
characteristic
|
parameter
|
|
range
|
0 to 9 999 999 pulse
|
|
Input impedance
|
>1Mohm
|
|
Trigger level
|
-1~14V in 1V steps and open collector inputs
|
|
Mini signal amplitude
|
2Vpp(pulse length>50us),3Vpp(pulse length 10..50us)
|
Switch test
|
characteristic
|
parameter
|
|
|
zero potential point
|
testing voltage(electric current)
|
2.8V(0.13A)perhaps24V(35A)
|
|
level detect
|
Input impedance
|
>1Mohm
|
|
|
Trigger level
|
-1~14V in 1V
|
2Generation, measurement, and simulation of electrical signals —— MC2-MFPortable multifunctional calibrator
millivoltmVMeasurement(T/CTerminal)-25~150mV
|
range
|
resolution
|
1Annual uncertainty
|
|
-25~150mV
|
0.001mV
|
0.02%RDG+4uV
|
millivoltmVGenerate(T/CTerminal)-25~150mV
|
range
|
resolution
|
1Annual uncertainty
|
|
-25~150mV
|
0.001mV
|
0.02%RDG+4uV
|
Voltage generation(-3~12V)
|
range
|
resolution
|
1Annual uncertainty
|
|
±0.25V
|
0.01mV
|
0.02%RDG+0.1mV
|
|
-3~0.25V
|
0.1mV
|
0.02%RDG+0.1mV
|
|
0.25~12V
|
0.1mV
|
0.02%RDG+0.1mV
|
milliamperemAmeasure(Source and suction current) 0~25mA
|
range
|
resolution
|
1Annual uncertainty
|
|
0~25mA
|
0.0001mA
|
0.02%RDG+1.5uA
|
Resistance measurement(0~4000 ohm(Ohms)
|
range
|
resolution
|
1Annual uncertainty
|
|
0~250(Ohms)
|
1mohm(Ohms)
|
4Wire connection:0.02%RDG+3.5mohm
3Wire connection:0.02%RDG+13.5mohm
|
|
250~2650 ohm(Ohms)
|
10mohm(Ohms)
|
|
2650~4000 ohm(Ohms)
|
100mohm(Ohms)
|
Resistance simulation(0~4000ohm(Ohms)
|
range
|
resolution
|
1Annual uncertainty
|
|
0~400ohm(ohm)
|
10mohm(ohm)
|
0.02%RDGperhaps30mohm(Choose the larger value)
|
|
400~4000ohm(ohm)
|
10mohm(ohm)
|
0.02%RDGperhaps30mohm(Choose the larger value)
|
Frequency generation(0.0005~10,000Hz)
|
range
|
resolution
|
1Annual uncertainty
|
|
0.0005~0.5Hz
|
0.000001 Hz
|
0.01%RDG
|
|
0.5~5 Hz
|
0.00001 Hz
|
0.01%RDG
|
|
5~50 Hz
|
0.0001 Hz
|
0.01%RDG
|
|
50~500 Hz
|
0.001 Hz
|
0.01%RDG
|
|
500~5000 Hz
|
0.01 Hz
|
0.01%RDG
|
|
5000~10000 Hz
|
0.1 Hz
|
0.01%RDG
|
Pulse counting
|
characteristic
|
parameter
|
|
range
|
0 to 9 999 999 pulse
|
|
resolution
|
1 pulse
|
|
Max load current
|
5mA
|
|
Output square wave amplitude
|
0~12Vpp±(0.2V+5%)
|
|
Output symmetrical wave amplitude
|
0~6Vpp±(0.2V+5%)
|
|
pulse frequency
|
0.0005~1000Hz
|
3Temperature signal measurement and simulation —— MC2-MFPortable multifunctional calibrator
thermocoupleT/CMeasurement and simulation
|
Standard thermocouple model
|
|
Graduation(2
|
range(℃)
|
range(℃)
|
1 Annual uncertainty(±)
|
B
|
0 ~ 1820
|
0~200
200~400
400~1820
|
(3
2.0 ℃
1.0 ℃
|
|
R
|
-50 ~ 1768
|
-50~0
0~100
100~1768
|
1.0 ℃
0.8 ℃
0.6 ℃
|
|
S
|
-50 ~ 1768
|
-50~0
0~1768
|
1.0 ℃
0.7 ℃
|
|
E
|
-270 ~ 1000
|
-270~-200
-200~1000
|
(3
0.25 ℃
|
|
J
|
-210 ~ 1200
|
-210~1200
|
0.3 ℃
|
|
K
|
-270 ~ 1372
|
-270~-200
-200~1000
1000~1372
|
(3
0.3 ℃
0.4 ℃
|
|
N
|
-270 ~ 1300
|
-270~-200
-200~1300
|
(3
0.4 ℃
|
|
T
|
-270 ~ 400
|
-270~-200
-200~-100
-100~400
|
(3
0.3 ℃
0.2 ℃
|
|
U
|
-200 ~ 600
|
-200~-100
-100~600
|
0.3 ℃
0.2 ℃
|
|
L
|
-200 ~ 900
|
-200~ 900
|
0.25℃
|
|
C
|
0 ~ 2315
|
0 ~ 1000
1000 ~ 2000
2000 ~ 2315
|
0.4 ℃
0.8 ℃
1.2 ℃
|
|
G
|
0 ~ 2315
|
0 ~ 100
100 ~ 2315
|
(3
1.0 ℃
|
|
D
|
0 ~ 2315
|
0 ~ 1000
1000 ~ 2000
2000 ~ 2315
|
0.4 ℃
0.8 ℃
1.2 ℃
|
thermal resistanceRTDMeasurement and simulation
|
sensor type
|
range
|
resolution
|
measure1Year uncertain(±)
|
simulate1Annual uncertainty(±)
|
|
Pt 50 ~ 1000
|
-200 ~ 200 ℃
200 ~ 600 ℃
600 ~ 850 ℃
|
0.01 ℃
0.01 ℃
0.01 ℃
|
0.1 ℃
0.2 ℃
0.3 ℃
|
0.15 ℃
0.25 ℃
0.35 ℃
|
|
Ni 100
|
-60 ~ 180 ℃
|
0.01 ℃
|
0.1 ℃
|
0.15 ℃
|
|
Ni 120
|
-80 ~ 260 ℃
|
0.01 ℃
|
0.1 ℃
|
0.15 ℃
|
|
Cu10
|
-200 ~ 260 ℃
|
0.01 ℃
|
0.2 ℃
|
0.8 ℃
|
|
Optional standardsRTDtype
|
|
Pt50 (385)
|
Pt400 (385)
|
Pt100 (3926)
|
Pt100 (3923)
|
Cu10 (427)
|
|
Pt100 (385)
|
Pt500 (385)
|
Pt100 (391)
|
Ni100 (618)
|
|
Pt200 (385)
|
Pt1000 (385)
|
Pt100 (375)
|
Ni120 (672)
|
thermocoupleT/Cmeasure/Simulated reference compensation
|
range(℃)
|
1Annual uncertainty
|
|
-10~50℃
|
±0.25 ℃
|
notes:
1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability, and long-term stability.Uncertainty does not include reference point uncertainty.
2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3) ±0.02 % of thermovoltage + 4 μV
4) DIN 43710
5) ASTM E 988 – 96
6) ASTM E 1751 - 95e1
4Pressure measurement and calibration module —— MC2-MFPortable multifunctional calibrator
high-precisionInternal pressure module(IPM) —— Installed inMC2-MFInside the portable multifunctional calibrator host:
|
Internal Module
|
unit
|
range
|
resolution
|
1Annual uncertainty
|
|
IPM200mC
|
kPa
|
±20
|
0.001
|
0.05%FS
|
|
IPM2C
|
kPa
|
-100~200
|
0.01
|
0.03%FS
|
|
IPM20C
|
kPa
|
-100~2000
|
0.1
|
0.03%FS
|
|
IPM160
|
MPa
|
0~16
|
0.001
|
0.03%FS
|
|
Atmospheric reference module
|
|
Absolute pressure module
|
|
Enable the instrument to measure both absolute pressure and gauge pressure simultaneously
|
high-precisionExternal pressure module(EXT)—— Installed inMC2-MFPortable multifunctional calibrator host external:
|
module
|
measure Cheng
|
refined linear measure
|
|
Absolute pressureBarometric
|
800~1200mbar abs
|
80~120 kPaAbsolute pressure
|
0.03kPa
|
|
EXT10mD
|
±10mbar differential
|
-1~1 kPa
|
0.05% Span
|
|
EXT100m
|
0~100mbar gauge
|
0~10 kPa
|
0.025% FS
|
|
EXT400Mc
|
±400bar
|
-40~40 kPa
|
0.025% FS
|
|
EXT1C
|
±1bar
|
-100~100 kPa
|
0.02% FS
|
|
EXT2C
|
-1~2bar
|
-100~200 kPa
|
0.015% FS
|
|
EXT6C
|
-1~6bar
|
-100~600 kPa
|
0.015% FS
|
|
EXT20C
|
-1~20bar
|
-0.1~2 MPa
|
0.015% FS
|
|
EXT60
|
0~60bar
|
0~6 MPa
|
0.02% FS
|
|
EXT100
|
0~100bar
|
0~10 MPa
|
0.02% FS
|
|
EXT160
|
0~160bar
|
0~16 MPa
|
0.02% FS
|
|
EXT250
|
0~250bar
|
0~25 MPa
|
0.02% FS
|
|
EXT600
|
0~600bar
|
0~60 MPa
|
0.02% FS
|
|
EXT1000
|
0~1000bar
|
0~100 MPa
|
0.02% FS
|
Conversion of pressure units:1bar = 0.1 MPa ;1bar = 100 kPa