The portable process signal calibrator MC2-TE, also known as the portable temperature calibrator, is a highly practical portable on-site thermal calibration tool. It is a product of Finland's Bemex company and has multiple functions for temperature and process signal measurement and calibration. Multiple external pressure modules can be selected to measure pressure and expand the pressure range; Measurement and simulation: Measurement and simulation of resistance, thermal resistance RTD, thermocouple TC, voltage, millivolt mV, current, pulse, and frequency, switch testing, HART compatibility, supplying 24V loop power. Chinese menu operation interface. Large screen, dual windows.
MC2-TE Portable process signal calibrator
|
Portable process signal calibrator Multi functional for measuring and verifying process signals, temperature, and electrical signals:
u thermal resistanceRTDmeasure/simulategenerate
u thermocoupleTCmeasure/simulategenerate
u resistance measurement/simulategenerate
u millivoltmVmeasure/simulategenerate
u Voltage generation
u Current generation (internal)/External power supply)
u Pulse generation
u Frequency generation
u Current measurement (internal)/External power supply)
u voltage measurement
u frequency
u PULSE COUNT
u internalHARTcompatible
u supply24VLoop Power
u Switch test
u You can choose to connect an external pressure module asPressure calibrator
u Chinese menu operation interface
|
High precision
Easy to use
large screen
Dual window display
Chinese Menu
Rechargeable Battery
Exquisite, portable, and durable
|
MC2-TE portable process signal calibratorTechnical parameter indicators:
|
characteristic
|
describe
|
|
display
|
60mm×60mm;LCDdisplay
|
|
weight
|
770g
|
|
size
|
215mm×102mm×49mm(d/w/h)
|
|
keyboard
|
membrane keyboard
|
|
Type
|
rechargeableNiMH,4000mAh,3.6V DC
|
|
charging time
|
5hour
|
|
Charging power supply
|
100~240VAC,50~60Hz
|
|
Battery settings
|
All day measurement mode, background light off
Half day mode provided12mACircuit power supply, background light on
|
|
Optional dry battery settings
|
8Hour measurement mode, background light off
4Hour measurement mode provided12mALoop Power,Background light on
|
|
operation temperature
|
-10~50℃(14~122℉)
|
|
storage temperature
|
-20~60℃(-4~140℉)
|
|
humidity
|
0arrive80%non-condensing
|
|
Max input voltage
|
30V DC,60V DC
|
|
safety standards
|
Derective 73/23/EEC,EN61010-1
|
|
EMC
|
Derective 89/336/EEC,EN61010-1
|
1Electrical signal measurement——MC2-TE portable process signal calibrator
voltage measurement(-1~60V DC)
|
range
|
resolution
|
accuracy
|
|
±0.25V
|
0.001mV
|
0.02%RDG+5uV
|
|
±(0.25~1V)
|
0.01mV
|
0.02%RDG+5uV
|
|
1~25V
|
0.1mV
|
0.02%RDG+0.25mV
|
|
25~60V
|
1mV
|
0.02%RDG+0.25mV
|
Milliampere measurement(100mA)
|
range
|
resolution
|
accuracy
|
|
±25mA
|
0.0001mA
|
0.02%RDG+1.5uA
|
|
±(25~100mA)
|
0.001mA
|
0.02%RDG+1.5uA
|
Loop Power
|
characteristic
|
parameter
|
|
output current
|
Max25mA,There is short circuit protection
|
|
Output voltage
|
24±10%
|
|
HARTOutput impedance in compatibility mode
|
300ohm±20%
|
frequency(0.0027~50,000Hz)
|
range
|
resolution
|
accuracy
|
|
0.0027~0.5Hz
|
0.000001 Hz
|
0.01%RDG
|
|
0.5~5 Hz
|
0.00001 Hz
|
0.01%RDG
|
|
5~50 Hz
|
0.0001 Hz
|
0.01%RDG
|
|
50~500 Hz
|
0.001 Hz
|
0.01%RDG
|
|
500~5000 Hz
|
0.01 Hz
|
0.01%RDG
|
|
5000~50000 Hz
|
0.1 Hz
|
0.01%RDG
|
Pulse counting
|
characteristic
|
parameter
|
|
range
|
0 to 9 999 999 pulse
|
|
Input impedance
|
>1Mohm
|
|
Trigger level
|
-1~14V in 1V steps and open collector inputs
|
|
Mini signal amplitude
|
2Vpp(pulse length>50us),3Vpp(pulse length 10..50us)
|
Switch test.
|
characteristic
|
parameter
|
|
|
zero potential point
|
testing voltage(electric current)
|
2.8V(0.13A)perhaps24V(35A)
|
|
level detect
|
Input impedance
|
>1Mohm
|
|
|
Trigger level
|
-1~14V in 1V
|
2Generation, measurement, and simulation of electrical signals——MC2-TE portable process signal calibrator
millivoltmVMeasurement(T/CTerminal)-25~150mV
|
range
|
resolution
|
accuracy
|
|
-25~150mV
|
0.001mV
|
0.02%RDG+4uV
|
millivoltmVGenerate(T/CTerminal)-25~150mV
|
range
|
resolution
|
accuracy
|
|
-25~150mV
|
0.001mV
|
0.02%RDG+4uV
|
Voltage generation(-3~12V)
|
range
|
resolution
|
accuracy
|
|
±0.25V
|
0.01mV
|
0.02%RDG+0.1mV
|
|
-3~0.25V
|
0.1mV
|
0.02%RDG+0.1mV
|
|
0.25~12V
|
0.1mV
|
0.02%RDG+0.1mV
|
milliamperemAmeasure(source/Inhalation current) 0~25mA
|
range
|
resolution
|
accuracy
|
|
0~25mA
|
0.0001mA
|
0.02%RDG+1.5uA
|
Resistance measurement(0~4000ohm)
|
range
|
resolution
|
accuracy
|
|
0~250ohm
|
1mohm
|
4Wire connection:0.02%RDG+3.5mohm
3Wire connection:0.02%RDG+13.5mohm
|
|
250~2650ohm
|
10mohm
|
|
2650~4000ohm
|
100mohm
|
Resistance simulation(0~4000ohm)
|
range
|
resolution
|
accuracy
|
|
0~400ohm
|
10mohm
|
0.02%RDGperhaps30mohm(Choose the larger one)
|
|
400~4000ohm
|
10mohm
|
0.02%RDGperhaps30mohm(Choose the larger one)
|
Frequency generation(0.0005~10,000Hz)
|
range
|
resolution
|
accuracy
|
|
0.0005~0.5Hz
|
0.000001 Hz
|
0.01%RDG
|
|
0.5~5 Hz
|
0.00001 Hz
|
0.01%RDG
|
|
5~50 Hz
|
0.0001 Hz
|
0.01%RDG
|
|
50~500 Hz
|
0.001 Hz
|
0.01%RDG
|
|
500~5000 Hz
|
0.01 Hz
|
0.01%RDG
|
|
5000~10000 Hz
|
0.1 Hz
|
0.01%RDG
|
Pulse counting
|
characteristic
|
parameter
|
|
range
|
0 to 9 999 999 pulse
|
|
resolution
|
1 pulse
|
|
Max load current
|
5mA
|
|
Output square wave amplitude
|
0~12Vpp±(0.2V+5%)
|
|
Output symmetrical wave amplitude
|
0~6Vpp±(0.2V+5%)
|
|
pulse frequency
|
0.0005~1000Hz
|
3Temperature signal measurement and simulation——MC2-TE portable process signal calibrator
thermocoupleT/CMeasurement and simulation
|
Standard thermocouple model
|
|
Graduation
|
range(℃)
|
range(℃)
|
Accuracy (±)
|
|
B (2
|
0 ~ 1820
|
0~200
200~400
400~1820
|
(3
2.0 ℃
1.0 ℃
|
|
R (2
|
-50 ~ 1768
|
-50~0
0~100
100~1768
|
1.0 ℃
0.8 ℃
0.6 ℃
|
|
S (2
|
-50 ~ 1768
|
-50~0
0~1768
|
1.0 ℃
0.7 ℃
|
|
E (2
|
-270 ~ 1000
|
-270~-200
-200~1000
|
(3
0.25 ℃
|
|
J (2
|
-210 ~ 1200
|
-210~1200
|
0.3 ℃
|
|
K (2
|
-270 ~ 1372
|
-270~-200
-200~1000
1000~1372
|
(3
0.3 ℃
0.4 ℃
|
|
N (2
|
-270 ~ 1300
|
-270~-200
-200~1300
|
(3
0.4 ℃
|
|
T (2
|
-270 ~ 400
|
-270~-200
-200~-100
-100~400
|
(3
0.3 ℃
0.2 ℃
|
|
U (4
|
-200 ~ 600
|
-200~-100
-100~600
|
0.3 ℃
0.2 ℃
|
|
L (4
|
-200 ~ 900
|
-200~ 900
|
0.25℃
|
|
C (5
|
0 ~ 2315
|
0 ~ 1000
1000 ~ 2000
2000 ~ 2315
|
0.4 ℃
0.8 ℃
1.2 ℃
|
|
G (6
|
0 ~ 2315
|
0 ~ 100
100 ~ 2315
|
(3
1.0 ℃
|
|
D (5
|
0 ~ 2315
|
0 ~ 1000
1000 ~ 2000
2000 ~ 2315
|
0.4 ℃
0.8 ℃
1.2 ℃
|
thermal resistanceRTDMeasurement and simulation
|
sensor type
|
range
|
resolution
|
measurement accuracy(±)
|
simulateaccuracy(±)
|
|
Pt 50 ~ 1000
|
-200 ~ 200 ℃
200 ~ 600 ℃
600 ~ 850 ℃
|
0.01 ℃
0.01 ℃
0.01 ℃
|
0.1 ℃
0.2 ℃
0.3 ℃
|
0.15 ℃
0.25 ℃
0.35 ℃
|
|
Ni 100
|
-60 ~ 180 ℃
|
0.01 ℃
|
0.1 ℃
|
0.15 ℃
|
|
Ni 120
|
-80 ~ 260 ℃
|
0.01 ℃
|
0.1 ℃
|
0.15 ℃
|
|
Cu10
|
-200 ~ 260 ℃
|
0.01 ℃
|
0.2 ℃
|
0.8 ℃
|
|
Optional standardsRTDtype
|
|
Pt50 (385)
|
Pt400 (385)
|
Pt100 (3926)
|
Pt100 (3923)
|
Cu10 (427)
|
|
Pt100 (385)
|
Pt500 (385)
|
Pt100 (391)
|
Ni100 (618)
|
|
Pt200 (385)
|
Pt1000 (385)
|
Pt100 (375)
|
Ni120 (672)
|
thermocoupleT/Cmeasure/Simulated reference compensation
|
range(℃)
|
accuracy
|
|
-10~50℃
|
±0.25 ℃
|
Portable process signal calibrator MC2-TE
For more detailed information on portable temperature calibrators, please call Wuhan Technical Service Center Welcome to request for free.