E-mail
2442648961@qq.com
-
Phone
13023169219
-
Address
No. 3666 Changcheng Road, Mudan District, Heze City, Shandong Province
Shandong Sairuite Testing Instrument Co., Ltd
2442648961@qq.com
13023169219
No. 3666 Changcheng Road, Mudan District, Heze City, Shandong Province
Quality assurance of reciprocating electrochemical corrosion friction and wear testerQuality assurance of reciprocating electrochemical corrosion friction and wear tester
Instrument Introduction:
The reciprocating electrochemical corrosion friction and wear tester is a specialized equipment for studying the corrosion wear and interaction mechanism of metal materials in electrochemical corrosion. The experimental process is automatically controlled by a computer to detect parameters such as test load, sliding speed, sliding length, friction coefficient, medium temperature, and electrochemistry, and displayed graphically and stored data. Accurately evaluate the abrasion characteristics of materials or components in a targeted manner.
Compliant with standards:
ASTM G133 "Standard Test Method for Sliding Wear of Linear Reciprocating Balls on Flat Plates"
Main parameters:
Control system: PLC or Windows
Operating interface: 22 inch monitor
Loading method: weight loading, automatic loading
Load range (standard):
Auto loading: 0-300N Accuracy: ± 0.25N
Weight loading: 500-2000g
Friction range (standard): 0-100N Accuracy: ± 0.25N
Reciprocating frequency: 0.05-20Hz
Repetition frequency: unlimited
Lifting height: 20mm
Journey: 2.5-40mm
Sample size: Φ 53 ± 0.5mm, thickness 1-5mm
Electrochemical testing temperature: room temperature to 40 ℃
Friction fixture: Φ 3, Φ 4, Φ 5, Φ 6mm
Friction pair: GCr15 steel ball, stainless steel ball, AlO and other ceramic balls
Electrochemical corrosion component bench
Storage tank size: 120mmx90mmx50mm, capacity 300m
1 set of electrochemical workstation
Power supply: AC220V 1.5KW
Electrochemical station parameters
Constant potential control range: ± 10V |
Constant current control range: ± 2.0A |
Potential control accuracy: 0.1% * full-scale reading ± 1mV |
Current control accuracy: 0.1% * full-scale reading |
|
Potential sensitivity: 10uV (>100Hz), 3 μ V (<10Hz)
|
Current sensitivity:<10pA |
|
Potential rise time:<1 μ S (<10uS) (<10mA), (<2A)
|
Current range: 2A~200nA, 8 levels in total |
Reference electrical input impedance: 1012 Ω | | 20pF |
Z large output current: 2.0A |
Slot pressure output: ± 21V |
Current scanning increment: 1mA @ 1A/mS |
CV and LSV scanning speed: 0.01mV~10000V/s |
DPV and NPV pulse width: 0.0001-1000s |
CA and CC pulse width: 0.0001~65000s |
Z small potential increment of CV: 0.075mV |
SWV frequency: 0.001~100KHZ |
Current and potential range: automatic setting |
|
AD data collection: 16bit@1MHz , 20bit @1kHz |
Low pass filter: 8-segment programmable |
DA resolution: 16 bits, setup time: 1 μ S |
|
Communication interface: USB 2.0 |
Basic configuration:
Test host 1 unit
Instruction manual: 1 copy
Certificate of conformity: 1 copy
Receipt: 1 copy
Wrench: 1 piece
Computer host: 1 set
Monitor: 1 set:
Communication cable: 1 set