E-mail
13622642640@163.com
-
Phone
13622642640
-
Address
No. 13 Jinyuling Road, Dongcheng Street, Dongguan City, Guangdong Province
Guangdong Innovation Instrument Technology Co., Ltd
13622642640@163.com
13622642640
No. 13 Jinyuling Road, Dongcheng Street, Dongguan City, Guangdong Province
Product Name and Model Index Introduction:
1.1working conditions:
●Operating Temperature15-30℃ ●Relative humidity:40%~50% ●Power supply:AC :220V ±5V
1.2. Technical performance and indicators:
1.2.1. The scope of elemental analysis ranges fromsulfur(S)To uranium(U);
1.2.2. The range of element content analysis is1PPmto99.99%;
1.2.3. Measurement time:100300Seconds;
1.2.4. RoHSinstructionstipulatedHarmful elements (limited)Cd/Pb/Cr/Hg/Br)Its detection limit is as high as1PPM;
1.2.5. The energy resolution is160±5electron volt;
1.2.6. The temperature adaptation range is15℃to30℃;
1.2.7 Power supply: AC220V±5V(It is recommended to configure an AC purification stabilized power supply.)
1.3.Product Features
1.3.1 EDX1800It is specifically targeted towardsROHSTheEN71Design a product in accordance with environmental regulations.
1.3.2Breaking the traditional straight line design of instruments, adopting a streamlined integrated design, the instrument is stylish and elegant.
1.3.3Adopting the United StatesAmptekdetectorSi-pinElectric refrigeration instead of liquid nitrogen refrigeration has the advantages of small size, accurate data analysis, and low maintenance cost.
1.3.4Adopting self-developed technologySESThe signal processing system effectively improves the sensitivity of measurement and makes the measurement more accurate.
1.3.5One click automatic testing, easier to use, more convenient, and more user-friendly.
1.3.6Seven types of optical path calibration and collimation systems, according todifferent samplesAutomatic switching.
1.3.7Multiple anti radiation leakage design, radiation protection level belongs to similar productsMedium to high level.
1.3.8The advanced integrated heat dissipation design greatly improves the overall heat dissipation performance and ensures the safe operation of core components.
1.3.9Temperature control technology for the movement ensuresXThe safe and reliable operation of the radiation source effectively extends its service life and reduces usage costs.
1.3.10Multiple instrument accessory protection system, and can be monitored throughout the entire process through software,Make the instrument work more stably and safely.
1.3.11ROHSSpecialized testing software, standard window design,friendly interface,operationConvenient.
1.3.12This machine adoptsUSB2.0The interface effectively ensures accurate, high-speed, and efficient transmission of data.
2.Product Configuration Checklist
Detailed configuration information:
1、 HardwareOne host, including the following main components:
(1)high powerXfluorescent tube (2) Electric cooled semiconductor detector
(3) Amplifying circuit (4) sample stage
(5) High definition camera (6) Automatically switch collimators
(7) Automatically switch filter (8) Signal to Noise Ratio Enhancer
(9) high-pressure system (10)Ultra large sample cavity
(11)Lenovo computer(12)HPprinter
2、 Software and others:
XFluorescence spectrometer full element analysis software,Test type (required): Hardware Plastic, etc
2.1One copy of the instrument user manual (including software operation manual and hardware operation manual), factory inspection certificate, packing list, warranty certificate, and other required documents.
2.2Standard Appendix
(1)Collimator hole: including Ø8.0mmF6.0mmF4.0mmF3.0mmF2.0mmF1.0mmF0.5mmF0.2mmOne each, in total8a. (Built into the instrument);
(2) Filter:5Combination of species (already built into the instrument);
(3) Standard sample (EU standard sample):PEStandard samplesA silver corrected film.
3. Specialized softwareJPSPEC-FP
4.1.1Software Introduction
specifically targeted atROHSDeveloped for detection, data processing, calculation, and reporting of collected spectral signals to display measurement results.
4.1.2Feature Introduction
※ Specially designed to deal withEuropean UnionRoHSIn the instructionSix substances involvedofsixSeed elementsCd, Pb, Hg, Br, Cr,Cltest,measurementtimefor100-300second
The operation interface is simple and intuitive, easy to use, and does not require professional operation※ Suitable calibration curves can be automatically selected for easier and more accurate measurement.
The interface automatically switches between Chinese and English, and has a third-party language customization function
※Automatic calibration instrument.
※ Comes with sample material qualitative analysis to prevent manual users from selecting the wrong curve
Multiple report formats are available for printing.
※ Multiple spectrograms can be displayed simultaneously
※Movement temperature monitoring technology ensuresXThe safe and reliable operation of the radiation source effectively extends its service life and reduces usage costs.