The FN-2000B pyroelectric radiation dose detection system can measure the pyroelectric detector after being irradiated by X-rays, gamma rays, beta rays, neutrons, and other radiation, and provide readout values and dose values. It has functions such as automatic calibration, background subtraction, heating parameter setting, and printing. The thermoluminescence dose reader can be connected to a computer, and data management, printing, and other functions can be performed through the application software of the thermoluminescence dose reader. Suitable for measuring from environmental background to radiation therapy dose, radiation flaw detection irradiation dose control, nuclear accident dose measurement monitoring, etc.
Performance characteristics:
7-inch touch screen with full Chinese color display, easy for users to operate, run, and maintain;
The high-voltage section has three modes: automatic, locked, and fixed;
The heating plate is replaceable, making it easy to measure different shapes and specifications of pyroelectric detectors;
The database supports unit name, individual name, gender, dose box barcode
Multiple query functions such as measurement date, tested personnel, and untested personnel;
• Power on automatic drying function, used for instruments with humid environments and low usage frequency;
An attenuation filter can be placed to expand the range of dose measurement.
Equipped with PC software, printer interface, barcode scanner interface; Single machine storage 1000
Measurement data.
Support scanning code input, table export, and personal database archiving.

Main performance indicators
| Optical measurement system | heating system | ||
| Dose measurement range | LiF:Mg,Ti:10-5Gy ~ 1Gy LiF:Mg,Cu,P:10-7Gy ~ 10Gy LiF:Mg,Ti-M:0~100Gy | Heating temperature range | Room temperature~400 ℃ |
| measurement items | X γ、β、 neutron | Repeatability of heating temperature | ≤1% |
| system stability | ≤0.2% | Heating temperature deviation | ≤±2℃ |
| Test reading time | 40s (customizable) | heating time | 0~500s |
| Change coefficient of sensitivity repeatability | ≤0.1%±0.05%/℃ | Repeatability of heating time | ≤0.1% |
| Drawer light leakage | ≤ 0.1% of the internal light source quantity | Heating rate | 1~40℃·s-1 |
| background value | ≤300 | Terms of Use | |
| high pressure | 0~1500V | working environment | indoor |
| Power supply | AC 220 V±10%,50Hz | Operating Temperature | 10~40℃ |