Overview
The GR-200 series thermoluminescence neutron dosimeter belongs to passive solid-state integral detectors. Its characteristics are small size (high spatial resolution), effective atomic number close to air and human tissue, no interference from electromagnetic radiation, wide measurement range, low measurable lower limit, good reproducibility, and pulse radiation field that can measure high dose rates. They can be applied to personal monitoring of ionizing radiation, environmental monitoring, and specific radiation field measurements (including differential measurements of n and gamma mixed fields). It can also be applied to disciplines such as radiology, radiobiology, reactor engineering, archaeology, etc.

Main performance indicators
| Detection object | X. Gamma rays; Material: LiF: Mg, Cu, P or LiF: Mg, Ti optional |
| Detection object | Neutron radiation; Material: 6LiF: Mg, Cu, P or 6LiF: Mg, Ti optional |
| shape | Circular discs (with optional square discs, powders, and square bars) |
| Specifications | Diameter 4.5 × 0.80mm (other specifications can be customized) |
| Energy response | (30KeV~3MeV photons):<20% |
| Relative sensitivity | 62 times (TLD-100) |
| linear range | 100nGy~12Gy |
| Detection threshold | 0.1μGy |
| dispersion | ≤± 5%, 3%, 1% (optional) |
| Response to neutrons | The response to absorbed dose caused by heavy particles is LET dependent, which is beneficial for distinguishing the gamma component in the mixed field of n and gamma |
| stability | Storage at room temperature for one month shows no significant decline, while storage at 50 ℃ for one month results in a decline of approximately 3% |
| Illumination effect | No significant response to indoor scattered sunlight, incandescent lights, and fluorescent lamps; Responds to direct sunlight (ultraviolet light) |
| working environment | T<50%; RH<90% |