Three purpose constant temperature water tank (tank)

NegotiableUpdate on 10/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Three purpose constant temperature water tank (tank)

Product Details

Technical Specifications:

model

Temperature control range

Temperature control accuracy

temperature uniformity

heating power

working dimensions

power supply

HH-600

Room temperature -100 ℃

≤1℃

≤1℃

1500W

600×280×150

220

Product Features:
The outer shell of this product is made of high-quality cold-rolled steel plate and electrostatically sprayed. The inner liner, upper cover, and panel are all made of high-quality stainless steel plate, and the heating method is immersion electric heating tube heating. The temperature control adopts a high-precision temperature regulator, which can be adjusted freely within the range of use.

Product Features:
The outer shell of this product is made of cold-rolled steel plate and electrostatically sprayed. The inner liner, upper cover, and panel are all made of stainless steel plate, and the heating method is immersion electric heating tube heating. The temperature control adopts a high-precision temperature regulator, which can be adjusted freely within the range of use.
Usage and precautions:
1. The water tank should be placed flat on a fixed platform, and the power supply voltage must match the required voltage of the water tank. The power socket should use a three hole safety socket, and the grounding wire must be properly connected before use.
2. Before use, pour clean water into the tank. To shorten the heating time, hot water can also be injected.
3. Plug in the power, turn on the water tank switch, click the function key, adjust the ↑ or ↓ key, and select the desired working temperature.
Repair and maintenance:
The temperature control instrument is a key component for controlling temperature. It has been accurately calibrated before leaving the factory. If a control failure is found, in addition to checking the circuit, the control instrument should be inspected for any faults or sent back to the company for repair.

Exception handling:

serial number

phenomenon

Possible faults

Solution

1

Display normal but not heating up

The set value is lower than the actual measured temperature

View setting values

2

The heating light is on but not heating up

Heating tube broken or output relay damaged

View and exclude

3

Display LLL and flash

Measuring temperature below -0.5 ℃ or probe breakage

View and exclude

4

Display 99.9 and flash

The measured liquid temperature is higher than the upper limit of the instrument measurement

View and exclude

5

The displayed temperature is higher than the actual value

Probe water ingress

Check and dry the probe

6

Display normal temperature out of control

Control the short circuit or open circuit of the heating element's circuit

View and exclude

7

The temperature never rises

The advance setting is too large

Reduce lead time

8

The displayed data is dim and shaky

Low working voltage

Check the working voltage