Welcome Customer !

Membership

Help

Dongguan Haoen Testing Instrument Co., Ltd. (Wuhan Branch)
Custom manufacturer

Main Products:

ybzhan>Products

UV aging test box

NegotiableUpdate on 12/21
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
UV aging test box, also known as UV aging test box, UV aging test box, UV aging test box, UV aging test box, UV aging test box, UV aging test box.
Product Details

UV aging test boxPriceUV aging test boxManufacturerbrand

1、 Technical indicators

Model: HE-UVA-3T

1. Studio size: 1140 × 400 × 390 mm (W × D × H)

2. Dimensions: 1300 × 500 × 1470 mm (W × D × H)

3. Irradiance control system (radiometer):

l Irradiance range: 0.45-0.9W/m2 adjustable (extends lamp life)

4. UV lamp tube parameters:

4-1 Lamp tube model:UV-A(315~400nm)、UV-B(280~315nm)

(Standard configuration includes 8 UV-A lamps. If UV-B lamps are required, please inform us in advance before placing an order as their prices are different.)

4-2 Lamp Brand:ATLAS in the United States*fluorescent tube

Center distance of 4-3 lamps: 70mm

Center distance between test sample and lamp tube: 50 ± 3mm

4-5 irradiation intensity: output according to the set light intensity

5. Temperature and humidity indicators of the test chamber:

5-1 Temperature range: adjustable from room temperature+10 to 70 ℃

5-2 Temperature fluctuation: ± 3 ℃

5-3 Temperature resolution: 0.1 ℃

5-4 Humidity range: ≥ 90% RH

6. Blackboard thermometer

6-1 Measurement range: 45-80 ℃

6-2 tolerance: ± 3 ℃

7. Temperature control method: PID self-tuning temperature control method

8. Water tank requirement: The water depth should not exceed 25mm and there should be an automatic water supply controller

IIFunctional Area Introduction:

1Radiation system (simulated sunlight):

Light source: 8 imported ultraviolet fluorescent lamps (ATLAS, USA) with a rated power of 40W, available in UVA-340 and UVB-313 light sources for users to choose from.

2. Condensation system (simulated dew):

The QUV condensation system essentially involves the evaporation of water vapor confined within the experimental chamber by the sample plate. The air on the back of the sample plate lowers the temperature of the water vapor, causing it to condense on the sample plate.

3. Sprinkler system (simulating rainwater):

The QUV sprinkler system consists of a high-power water pump, 8 nozzles (4 on each side), connecting pipes, control valves, and drainage parts. The nozzle is installed between the lamp tubes, and in the spray cycle, water is sprayed onto the sample to produce thermal shock or erosion effects.

4. Suneye (radiometer) system:

Changes in light intensity can affect the rate of material aging, while differences in spectra may simultaneously affect both the rate and type of material aging. Accurate and reproducible test results can only be obtained by controlling irradiance through the solar eye system.

The solar eye system is a precise closed-loop control system that can automatically maintain the set light intensity through a feedback loop. The controller can continuously monitor the intensity of ultraviolet light and compensate for changes in light intensity caused by lamp aging or any other factors by adjusting the output of the lamp. The reproducibility and repeatability of the test results of the solar eye irradiance control system are better than those of manual irradiance control, and can effectively extend the life of the lamp tube.

Similar Product Recommend