2、 Technical parameters:
|
model |
SHA-C |
SHZ-82 |
SHA-B |
|
Temperature control range |
Room temperature -100 ℃ |
||
|
temperature control accuracy |
±0.5℃ |
||
|
temperature fluctuation |
±0.5℃ |
||
|
heating power |
1800W |
||
|
Oscillation mode |
Round trip |
Rotary type |
Dual function of round-trip/revolving |
|
Oscillation amplitude |
20mm |
||
|
speed regulation range |
40-300 revolutions per minute |
||
|
timing range |
0-120 minutes or normally open |
||
|
inner dimensions |
490×390×170mm |
||
|
Overall dimensions |
700×550×490mm |
700×550×520mm |
|
|
Bottled quantity |
Test tube diameter: 16 × 300 100ml × 24 200ml × 15 250ml × 12, 500ml × 6, 1000ml × 4 |
||
|
Total weight |
38kg |
40kg |
48kg |
|
power supply |
220v 50Hz±10% |
||
3、 Usage:
1. Add an appropriate amount of water to the water tank (the water level should be 2-3cm above the spring clamp) and place the reagent bottle on the spring clamp.
2. Firstly, turn on the power, turn on the heating switch, and the temperature controller will light up. At this time, click the SET function button on the temperature controller. Click the set function button on the temperature controller, and the upper row will display sp and the lower row will display the set temperature. Click the △ or ▽ button to select the desired working temperature. After setting, click the set function button again to enter the constant temperature timing setting. The upper row will display st and the lower row will display the set time. Click the ▲ or ▼ button to select the constant temperature timing time. (If no timing is required, set the time to 0) Click the set button again to exit the setting state. At this time, the temperature inside the box is running at a constant temperature according to the set temperature; After running, the buzzer will sound an alarm and stop heating.
3. Adjust the oscillation "timing knob", select the oscillation timing time (if timing is not required, adjust the timing knob to the normally open position), and then slowly adjust the "speed control knob" to the desired oscillation speed.
4. After use, adjust the speed control knob to the 0 position, turn off the power. If not in use for a long time, unplug the power, drain the water, clean and wipe it clean in a timely manner to extend its service life.
4、 Precautions for use:
1. The power socket provided by the user should have good grounding measures.
2. It is strictly prohibited to move the machine during normal operation.
3. It is strictly prohibited for objects to collide with the machine.
4. Children are strictly prohibited from approaching the machine to prevent accidents.
5. It is strictly prohibited to dry burn the water bath constant temperature oscillator without water during use.