Intelligent eddy current separatorFunctional Information
·Real time display of F1, F3, F5, F7 spectrograms
·Display F1, F2, F3, F4, F5, F6 or F7 impedance plan separately
·Eight independent hardware alarm output ports, providing external trigger sorting signals to achieve automatic sorting
·Can store various detection programs and results in file format for massive storage
·The material properties of interest can be directly displayed in numerical form (regression analysis)
·Can be paired with a matching pairDifferential and self comparison probes
·Menu based human-machine dialogue (in both Chinese and English versions, with optional Chinese complex/simple fonts)
·Equipped with configuration analysis function: Due to the mapping of electromagnetic characteristics of similar metal materials on the eddy current impedance plane following a two-dimensional normal distribution, the instrument automatically forms an elliptical sorting domain with a selection degree of 60.00% to 99.99% (adjustable) using probability statistical methods
·Using pre multi frequency technology to calibrate and classify components of different materials or heat treatment states, in order to improve the reliability of detection and the ability to distinguish different parameters (the instrument can preset up to 8 frequencies)
·Regression analysis function, such as when measuring hardness or hardened layer depth, the instrument can determine the regression function based on a set of known specimen values, calibrate the instrument, and directly display the test results in numbers, making it convenient for operators to perform non-destructive evaluation (NDE)
·The various operational parameters of EEC-41 can be stored in file format for calling during detection. The eddy current signal detection process can be recorded on a hard disk or optical disc
Intelligent eddy current separatorTechnical Specifications
Size and weight:
Size: 490 * 426 * 177mm Weight: 20kg
Power supply: AC 220V/110V
Working environment:
Working temperature: 0-40 ℃
Storage temperature: -20 ℃~+50 ℃
Relative humidity: not exceeding 85%
Eddy current testing:
1. Channel: 1
2. Frequency: 8
3. Frequency range: 5Hz~500KHz
4. Excitation waveform: sine wave
5. Magnification range: continuously adjustable from 1 to 256, in 1-step increments
6. Probe excitation signal amplitude: 1-8 levels
7. Pre amplification: 10-35dB
8. Standard sample preset: 8
9. Impedance interface: 8
Application field
Suitable for testing the matrix structure, surface hardness, hardening depth, strength, and material mixing of steel materials
Shanghai Jiuheng Instrument Co., Ltd. is an enterprise specializing in the research and development, production, sales, and service of instruments and meters. Ultrasonic flaw detectors: (EPOCH 600 ultrasonic flaw detector, EPOCH 6LT ultrasonic flaw detector, EPOCH 650 ultrasonic flaw detector, EPOCH 1000 ultrasonic flaw detector, USM 36 ultrasonic flaw detector, USN 60 ultrasonic flaw detector, USM go+ultrasonic flaw detector, Ominskan SX phased array flaw detector, OmniScan MX2 phased array TOFD flaw detector); Ultrasonic thickness gauge: (27MG ultrasonic thickness gauge, 45MG ultrasonic thickness gauge, 38DL PLUS ultrasonic thickness gauge, DM5E ultrasonic thickness gauge, CL5 ultrasonic thickness gauge, MAGNA-MIKE 8600 Hall effect thickness gauge, ETG-100 electromagnetic high-temperature thickness gauge); Ultrasonic probe; Customized phased array probe; Eddy current flaw detector: (NORTEC 600 eddy current flaw detector); Hardness tester; Endoscopes: (IPLEX NX Olympus Endoscope, IPLEX GX/GT Olympus Endoscope, IPLEX G Lite Olympus Endoscope, IPLEX TX Olympus Endoscope, IPLEX YS Olympus Endoscope, XLGo+Industrial Video Endoscope GE, XL Detect+Industrial Video Endoscope GE New, UVin Professional UV Video Endoscope); Pulse generator: (DRP300 replaces Olympus' 5072PR, 5073PR); Alloy analyzer: (Vanta Olympus handheld alloy analyzer, DELTA Professional handheld alloy analyzer Olympus, DELTA Element handheld alloy analyzer Olympus) and other instruments and meters. Shanghai Jiuheng Instrument Co., Ltd. SWC-2 transverse wave coupling agent, coupling agent for transverse wave probes.