GC-2060

NegotiableUpdate on 05/04
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
GC-2060 Main technical indicators: Temperature control indicators: ● Temperature control area: Chromatography column box
Product Details

Main technical indicators of GC-2060:
warmControl indicators:
● Temperature controlled area: chromatographic column box, samplerDetector, auxiliary device

controlwarmrangeroom temperature5℃400℃(With)1℃Incremental)

Control accuracy:Better than ±0.1
Program heating up:Program heating order: 5 orders Time setting: 99.9 (min)

Heating rate: 0.140 ℃/min (set in increments of 0.1 ℃, maximum heating rate below 200 ℃ can reach 40 ℃/min)


GC-2060 detector technical parameters:

Technical Parameter

Sensitivity/Detection Limit

Baseline noisesound

Baseline drift(30min

linear range

Hydrogen flame (FID)

≤2×10-11g/s(Hexadecane)

≤2.5×10-14A

≤5×10-14A

105

Thermal conductivity pool (TCD)

S≥2500mV•ml/mg(benzene/Toluene)

≤0.02mv

≤0.05mv

104

Flame luminosity (FPD)

(S)≤5×10-11g/smethyl-parathion
(
P)≤3×10-11g/smethyl-parathion

≤2×10-12A

≤1×10-12A

102
103